Membrane Output Calculation
Membranes produce the rated gallons per day (GPD) at 60 psi (4.1 bars)
operating pressure, 77°F (25°C) operating temperature and 500 ppm total
dissolved solids.
Membrane output gallons per day (GPD) depends on operating pressure,
water temperature and the ppm TDS in the feed water.
Expected GPD = Rated GPD × PCF × TCF
PCF
is the pressure correction factor
TCF
is the temperature correction factor
Calculation of Pressure Correction Factor (PCF):
The output (GPD) from
the membrane is directly proportional to the applied pressure.
The membrane is rated to produce the rated GPD at 60 psi. For any
pressure other than 60 psi the output GPD is multipled by the PCF.
PCF = Line Pressure (in psi) ÷ 60
Calculation of Temperature Correction Factor (TCF):
The output (GPD)
also decreases with decrease in temperature. This is because water viscosity
increases with decrease in water temperature.
Temperature Correction Factor Table (TCF)
°F /°C
TCF
°F /°C
TCF
°F /°C
TCF
41.0 /5
0.521
59.0 /15
0.730
77.0 /25
1.000
42.8 /6
0.540
60.8 /16
0.754
78.8 /26
1.031
44.6 /7
0.560
62.6 /17
0.779
80.6 /27
1.063
46.4 /8
0.578
64.4 /18
0.804
82.4 /28
1.094
48.2 /9
0.598
66.2 /19
0.830
84.2 /29
1.127
50.0 /10
0.620
68.0 /20
0.857
86.0 /30
1.161
51.8 /11
0.640
69.8 /21
0.884
87.8 /31
1.196
53.6 /12
0.661
71.6 /22
0.912
89.6 /32
1.232
55.4 /13
0.684
73.4 /23
0.941
91.4 /33
1.267
57.2/14
0.707
75.2 /24
0.970
93.2 /34
1.304