Chapter 4 Operation

4.1 Power on

Press the power button and hold for 2 seconds to enter into the interface as below:

4.2 Mode Selection

4.2.1 ETS

EMG Triggered  Stimulation  (ETS)  is  a  therapeutic  fusion  between  sEMG  Bio-
feedback  and  NMES.  The  stimulator  delivers  electrical  stimulation  according 
to  the EMG signal of the body part.  During ETS, the patient is required to  co-
ntract  the  muscle  as  much  as  possible  against  a  target  threshold.  Once  the 
target  is  achieved,  the  stimulation  is  triggered  to  against  a  target  threshold. 
Once  the  target  is  achieved,  the  stimulation  is  triggered  to  supplement  the 
naturally  achieved  muscle  contraction  beyond  the  contracting  muscle's  actual 
ability.  There  are  2  ways  to  achieve  a  threshold  of  stimulation:  1.  The  user 
manually downloads the threshold from the iPad APP or; 2. The stimulator dete-
cts  the  EMG  signal  and  sets  a  threshold  based  on  the  signal:  the  screen  will 
display “Relax” and then “Flex”, after which the threshold will be set.  

The default mode is ETS once the stimulator is powered on. The user can down-
load a threshold from the iPad APP or the stimulator will work on the default thres-
hold,  the  threshold  is  displayed  in  the  middle  of  the  yellow  area.  When  down-
loading 0 from the iPad APP, AUTO will be displayed in the middle of yellow area.  

4.3.4 Time Locking

The  treatment time  can  be  downloaded from  the APP  and  the  stimulator will 
operate within the specified time, once the prescribed time is up, the stimulator 
will shut down and cannot work again until downloading another treatment time 
from the APP.

4.3.5 Current Limit

A current limit value can be downloaded from the APP to limit the maximum output 
current. The available working current of the stimulator is 0 to 60 mA, however, if 
the user downloads a current limit value of 25 mA from the APP, then the maximum 
output current will stop at 25mA.

When  using  stimulation  modes  NMES  or  ETS.  As  a  safety  feature  the  user  must 
first  press  the  “ -“   button  twice  beforepressing  the  “ +”   key  to  increase  intensity. 
This is to ensure that an unwanted increase in “mA” intensity does not occur. The 
stimulator is designed to automatically lock after 10 seconds. To unlock and increase 
intensity the user must first press the “-“button twice and then proceed to pressing 
the “+” button to increase intensity to the desired setting.

4.1.1 Safety Feature

4.3.2 Loose electrode warning 

When the electrodes are not properly contacted with the skin, the output will 
stop and the screen will display a electrodes loose  warning. Please check the 
electrodes and reattach them, making full contact with the skin, then press        
to start again. The followings are the conditions that will cause a electrodes 
loose warning:

1.  In ETS or GAME mode, if any of the three electrodes are loose
2.  In NMES mode, if any of the yellow electrodes are loose

4.3.3 Data Storage

The data which can be stored in the system includes: current limit value, previ-
ously used prescription, previously downloaded threshold, the rest time for treat-
ment;  and  parameters  for  each  prescription  in  NMES  mode:  frequency,  pulse 
width, ramp up, ramp down, duration, interval, etc.

Summary of Contents for 01AS80

Page 1: ...m Date 2016 12 23 sEMG Biofeedback FES Neuro Muscular Electrical Stimulator 01AS80 User manual Thanks for buying our product Please carefully read the user manual before use and keep the manual in a safe place ...

Page 2: ...ions Skin Care Chapter 3 The Aspire 2 Description Features Settings Chapter 4 Operation Power on Mode Selection Other Functions Chapter 5 Care and Maintenance Stimulator Battery Lead Wires Guardian Self Adhesive Electrodes Chapter 6 Specifications and Parameters Chapter 7 Manufacturer s Declaration Warranty Customer Service 01 01 01 02 04 04 04 04 05 06 06 07 07 07 08 09 09 09 13 15 15 15 16 16 17...

Page 3: ...rehabilitation treatment programs for each patient 1 2 Main Innovations 1 Technology of self EMG signal acquisition and processing 2 New generation of real time biofeedback functional electrical stimulation technology based on EMG signal and muscle strength 3 iPad interactive games training and rehabilitation therapy applications APP 4 Multiple rehabilitation training modes in the APP Warranty Spe...

Page 4: ...It is customer responsibility to assure that this equipment and vicinity equipment complies the value of radio frequency inter ferences shown in General Regulation for safety according to IEC 60601 1 2 Tables as described in this section The manufacturer is not responsible for any interference caused by using other than recommended interconnect cables or by unauthorized changes or modifcations to ...

Page 5: ... 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequenc...

Page 6: ...ld strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol MANUFACTURER S DECLARATION ELECTROMAGNETIC IMMUNITY This Aspire 2 is intended for use in the electromagnetic environment specified below The customer or the u...

Page 7: ... 2 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment The Aspire 2 is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings and homes used for domestic purposes MANUFACTU...

Page 8: ...atio 100dB NMES Frequency 5 100Hz biphasic balanced wave 10 Pulse width 50 450uS 10 Output intensity load 1000Ω min 1mA max 60mA ETS Frequency 18 Hz biphasic balanced wave 10 Pulse width 150 uS 10 Output intensity load 1000Ω min 1mA max 60mA Working Environment Temperature 5 40 Relative humidity 80 HR Atmos 86Kpa 106Kpa Storage Environment Temperature 20 55 Relative humidity 93 HR Atmos 70Kpa 106K...

Page 9: ...wires carefully after each use Guardian Self Adhesive Electrodes Check the short connectors to make sure they have not become separated from the electrodes Replace electrodes onto plastic film after use If they drop onto the floor debris will adhere to conductive gel rendering the electrodes ineffective Optimal electrode function cannot be guaranteed after 4 uses Keep electrodes away from water Tr...

Page 10: ...o not use cleaning sprays or alcohol based cleaning solutions Do not store in direct sunlight in high temperatures moist areas dusty areas or near corrosive gas Do not immerse in water Do not throw step on or exert pressure on the stimulator When the stimulator reaches the end of its service life 3 years please dispose of it in accordance with the local and national regulations Battery The Aspire2...

Page 11: ... up the stimulator will shut down and cannot work again until downloading another treatment time from the APP 4 3 5 Current Limit A current limit value can be downloaded from the APP to limit the maximum output current The available working current of the stimulator is 0 to 60 mA however if the user downloads a current limit value of 25 mA from the APP then the maximum output current will stop at ...

Page 12: ...e effort exerted and gives them a tangible goal to work towards thereby improving their performance and affecting a physiological change The animations provide an enjoyable engaging format during the rehabilitation process Different games have been designed specifically to address 3 different types of muscle function strength endurance and coordina tion The system transfers the collected EMG signa...

Page 13: ...r muscle movements as the reliability of the signal will be directly impacted by the type of swallow elicited such as dry swallow 5cc water 10cc water puree or solid bolus Proper use of sEMG can be very useful in providing objective data for establishing baseline and progress over time Improper use of sEMG will not yield useful or accurate data The stimulator has 5 prescriptions as follows 4 fixed...

Reviews: