Spectradyne LLC, All rights reserved
37
Version 2.5.0 Revision 3, July 2020
Data Export
Navigate to the Export frame in the Data Output tab of the Viewer.
➢
Use the radio button to export all loaded files or the selected file only in the Currently
Loaded list.
Export Options
➢
Background Subtraction:
Select whether the estimated size distribution of electrical
noise false positives should be subtracted from the exported data. See page 32 for more
information about background subtraction.
➢
Binning:
Choose bin widths having a percentage of center value (default), or equal bin
sizes.
➢
Units:
❖
Number-weighted or volume-weighted distributions
❖
Concentration Density or Absolute Concentration
Run Report(s)
➢
Generate a 1-page Run Report summarizing the selected measurement.
➢
The Run Report is an Excel (.xlsx) format file containing a complete summary of the
measurement. It contains:
o
Meta information about the measurement.
o
Graphical size distributions in linear and logarithmic y axes.
o
Summary statistics.
➢
If multiple files are chosen, a separate Run Report will be generated for each file.
Export Size Distribution Data
Export the particle concentration vs. size distribution to external format.
➢
Click the “Size Distribution Data” button in the export frame.
➢
Specify the desired export options.
➢
Click “Export” and specify a file name and save location for the report.
Peaks Data
Export all peak data, including background subtracted peak data, to Excel (.xlsx) format.
➢
Click the “
Peaks Data
” button in the export frame.
Integration Range Report
Export the measured concentration for one or more user-specified integration ranges.
➢
Click the
“Integration Range Report”
button in the export frame.
➢
Specify the desired export options.
➢
Click “Export” and specify a file name and save location for the report.