Utilities
Instrument Prime
Run this sequence to prepare the instrument for running the first cartridge on startup. Follow the
prompts from the software:
➢
Load a reusable cleaning cartridge
➢
Allow the routine to complete:
o
Fill Monitor 1 and 3 sliders should reach the top.
➢
Lower the stage
➢
Remove the reusable cleaning cartridge and store it in a safe place
➢
Load the sample cartridge and run as usual
Clear Constriction
This automatic sequence is used during a measurement run to clear the sensing constriction of
the cartridge in the event of a blockage.
➢
Press
the “Clear Constriction” button
(Figure 9) and wait for the routine to complete.
➢
Continue acquisitions as normal
(press “Continuous” on Acquisition panel)
.
How much data is enough?
➢
The 1-
σ error reported for a concentration measurement in a given
size range is derived from statistical error in counting particle
detection events.
➢
Particle detection events are Poisson-distributed in time; therefore,
the error scales as 1/N
1/2
(with small corrections for N < 5).
➢
Example: N = 100 particles must be counted in a given size range to
obtain a concentration measurement on that range having standard
error of 10 %.
➢
The particle concentration (with error) measured over any user-
defined size range can be obtained in real time using the Live CSD
feature of the Auto-analysis engine.