![SPECS PHOIBOS 100 User Manual Download Page 8](http://html1.mh-extra.com/html/specs/phoibos-100/phoibos-100_user-manual_1343047008.webp)
Introduction
bus interface. All parts are integrated into a single, compact RF-shielded aluminum
case. The detection electronics are supplied together with the analyzer control unit.
PHOIBOS analyzers provide the detection of electron and ion energies between 0 and
3500 eV with minimum step widths of 7 meV. For ultra high energy resolution applica-
tions the unit can be operated in a 400 V or a 40 V range with extremely low ripple.
Step widths down to 80
μ
V are possible in the lowest voltage range.
A multi-element, two-stage transfer lens can be operated in several different modes for
angular and spatially resolved studies. All lens modes can be set electronically. A Slit-Or-
bit mechanism and a Multi-Mode Lens make the sampling area of the analyzer and the
acceptance angle area of the lens selectable. Thus the analyzer allows spatially resolved
measurements down to a diameter of 100
μ
m as well as large area investigations associ-
ated with different lens acceptance angles.
All units are completely controlled by SPECS software. Operation of the software will be
described in a separate manual.
Typical uses of the PHOIBOS analyzer include photoelectron spectroscopy (XPS, SSXPS,
UPS), Auger electron spectroscopy (AES, SAM) and ion scattering spectroscopy (ISS). The
PHOIBOS analyzer is bakeable up to 200°C after removal of the detector electronics and
the connection for the lens supply.
2
Summary of Contents for PHOIBOS 100
Page 1: ...PHOIBOS Hemispherical Energy Analyzer Series PHOIBOS 100 PHOIBOS 150 3 1...
Page 6: ...Table of Contents PHOIBOS...
Page 10: ...Introduction 4 PHOIBOS...
Page 13: ...Electrical Connections Figure 2 Connection Scheme PHOIBOS 7...
Page 14: ...Components and Connections Figure 3 Analyzer Housing PHOIBOS100 8 PHOIBOS...
Page 15: ...Electrical Connections Figure 4 Analyzer Housing PHOIBOS150 PHOIBOS 9...
Page 42: ......
Page 51: ...SpecsLab Hardware and Software Installation PHOIBOS 45...
Page 52: ......
Page 62: ......
Page 78: ...Analyzer Checks Figure 33 Schematics of the 12 pin Analyzer Feedthrough 72 PHOIBOS...
Page 80: ......
Page 86: ......
Page 92: ...List of Figures Figure 39 Alignment Pin 84 II PHOIBOS...
Page 93: ...List of Figures PHOIBOS III...
Page 94: ......
Page 96: ......
Page 98: ...Index PHOIBOS...