Possible Problems
procedure no.
→
incorrect detector preamplifier
threshold
→
high noise of the primary source
use a second excitation source,
electric interference
monitor mains voltage, check ground connec-
→
Table 12: Noisy Spectrum
possible cause
perform check, test, or troubleshooting
procedure no.
→
malfunction of lenses
Check tubus voltages section , ‘Check all ana-
lyzer (HSA 3500) voltages” on page . The
voltage can be measured either at the 12 pin
→
the rear of the HSA3500.
improper adjustment of the
Slit Orbit mechanism
check proper adjustment section
→
3.4 ,
section 4.4
magnetic fields
→
Table 13: Incorrect area was analyzed in all lens modes
55
Summary of Contents for PHOIBOS 100
Page 1: ...PHOIBOS Hemispherical Energy Analyzer Series PHOIBOS 100 PHOIBOS 150 3 1...
Page 6: ...Table of Contents PHOIBOS...
Page 10: ...Introduction 4 PHOIBOS...
Page 13: ...Electrical Connections Figure 2 Connection Scheme PHOIBOS 7...
Page 14: ...Components and Connections Figure 3 Analyzer Housing PHOIBOS100 8 PHOIBOS...
Page 15: ...Electrical Connections Figure 4 Analyzer Housing PHOIBOS150 PHOIBOS 9...
Page 42: ......
Page 51: ...SpecsLab Hardware and Software Installation PHOIBOS 45...
Page 52: ......
Page 62: ......
Page 78: ...Analyzer Checks Figure 33 Schematics of the 12 pin Analyzer Feedthrough 72 PHOIBOS...
Page 80: ......
Page 86: ......
Page 92: ...List of Figures Figure 39 Alignment Pin 84 II PHOIBOS...
Page 93: ...List of Figures PHOIBOS III...
Page 94: ......
Page 96: ......
Page 98: ...Index PHOIBOS...