Step3. 

Look into the eyepiece and turn the calibration screw till the boundary between the upper blue field 

and the lower white field meet exactly on the zero scale, such as 

shown in the image. That is the end of the calibration process. Make 

sure the ambient room temperature is correct for the solution you are 

using (20

/68

).  When working temperature of the room or 

environment (not the sample) changes by more than 5

,  we 

recommend recalibrating to maintain accuracy. If the instrument is 

equipped with Automatic Temperature Compensation system, the 

ambient working temperature of the room must be 

20

/68

whenever the instrument is recalibrated. Once calibrated, 

shifts in ambient temperature within the acceptable range (10

-30

should not affect accuracy

 

                                                                                                                                   Calibrate to “0” scale 

 

 

 

 

Step4. 

Now place a few drops of the sample to be tested onto the main 

prism, close the daylight plate and check reading. Take the reading 

where the boundary line of blue and white separating the graduated 

scale. The scale will provide a direct reading of the concentration. 

Each model of scale different, but the operation method is 

totally the same. 

 

 

 

 

                                                                                                                         Reading of Sample 

 

 

Warning-Maintenance 

 

1. Accurate measurement depends on careful calibration. The prism and sample must be at the same  

    temperature for accurate results. 

2. Don’t expose the instrument to damp working conditions, and don’t immerse the instrument in  

    water. If the instrument becomes foggy, water has entered the body. Call a qualified service 

    technician or contact your dealer. 

3. Don’t measure abrasive or corrosive chemicals with these instruments. They can damage the  

    prism’s coating. 

4. Clean the instrument after measurement using a soft damp cloth. Failure to clean the prism on a  

    regular basis will lead to inaccurate results and damage to the prism’s coating. 

5. This is an optical instrument. It needs careful handling and storage. Failure to do so can result in  

    damage to the optical components and its basis structure. With care, this instrument will last  

    years of reliable service. 

 

 