TURNING ON 

 
*   REMOTE switch turns to „OFF“ position 

 

- local control mode 

*   All potentiometers move to the left zero position 

 

- only advice, it’s not strong 

demand 
*   Disable OUTPUT switch move to „ON“ position 

- its not strong demand 

*   Connect a load 

 

 

 

 

 

- it’s not necessary 

*   Cable plug insert to a socket 
*   Switch on the mains switch – position „ON“ 
     (mains switch indicator and both LED displays light on) 
*   Set assumed current limitation by potentiometers 

 

- in CV mode 

     (ROUGHLY or FINE) 
*   Set desirable output voltage on the both displays  
     may be observed real date of the voltage or current 
*   Set assumed max. voltage depending on the max.  

- in CC mode 

     load resistance and max. constant current 
*   Set desirable constant current within constant       
     current range 
 
During operation in CV or CC mode an output power supply may be interrupted and restarted again 
by the disable OUTPUT switch. 
 

TURNING OFF 

 
It’s recommended to turn the load off, first by the disable OUTPUT switch and to remove wires from 
the output terminals. Than the mains switch may be turned off. Unit may be switched off directly by 
the mains switch at any operating condition.  
 

MAINTENANCE 

 
Power Supply SPS 1000 normally needs no maintenance or calibration. Only care must be taken not 
to obstruct cooling of the unit. When a unit is equipped by a dust filter, which must be cleaned or 
changed for a new one at the least once per year. The built-up of dust on the impellers of the fans and 
the heat sinks fins depends on the environment. Since the fans have over-capacity dust will not 
present a problem very quickly. The internal construction of the power supply is such that no dust 
wouldn’t reach the sensitive control circuit. There may occur a dust between the heat sink fins. It is 
advisable to inspect the fans, heat sinks as well as all units regularly. 
 

IMPORTANT ADVICE AND CAUTION   

 
- Electric strength (EDV) tests should not be repeated in the field. Improper test methods can cause 

severe damage to the power supply. Electric strength and earth continuity tests have been performed 
in the factory on each unit. The 4000V

rms

 cannot be tested afterwards on the assembled unit because 

the insulation between the components on the input side to the case is specified at 2000V

rms

. Since 

   the insulation output-case is low (only 500V DC due to fans) the insulation of the primary     
   components to case will break down when 4000V

rms

 is applied between input and output  

Summary of Contents for SPS1000 SERIES

Page 1: ...t is limited approximately to 60A 10ms This must be taken into account at a fusing of the line outlet It is recommended to use 16A normal fuse or 10A slow one The rectified 230V AC 300V DC is chopped...

Page 2: ...nit Voltage is indicated on the 3 digit display and the current at the same time on the 3 digit display Power supply output may be turned on or off by the Disable OUTPUT switch at any time independent...

Page 3: ...PE 50Hz 230V 10 TN S Protection achieved by basic insulation and connection to earth for all conductive parts that are capable of assuming hazardous voltage if the basic isolation fails An output sec...

Page 4: ...ondition MAINTENANCE Power Supply SPS 1000 normally needs no maintenance or calibration Only care must be taken not to obstruct cooling of the unit When a unit is equipped by a dust filter which must...

Page 5: ...oad inductance will not form a serial resonant circuit The output voltage is available 2s after mains switch on This feature may be improved according to customer requirements Serial operation is allo...

Page 6: ...TTL logic and lower one could be able to trigger the thyristor crowbar under normal operating condition due to a disturb noise Avoid condensation inside the power supply break down could be the resul...

Page 7: ...after this string send character 0x0D ENTER Meaning of each symbol S beginning of message Z power supply switched ON V power supply switched OFF Tu switch OFF power supply resp his output zero voltage...

Reviews: