S5 System • Appendix
7.6
SM-5811-0000.01
ENG
Shelf with AC outlet
System panel
3/4/5
position
Weight - shelf
approx. 6.5 kg
Maximum load - shelf
8 kg
Number of sockets
4
Protection
at 230/240 V: Circuit breaker 2 A
Protection
at 110/115 V: Circuit breaker 2 A
Load rating
2 A maximum in total
Sum of leakage currents
500 μA max. in total
Display and control module / touchscreen
Operating voltage
24 V
Power consumption
45 W
Pixel Failure Class
Conformity with Pixel Failure Class III
Level
Alarm limit (level sensor) for oxygenators/reservoirs made of
rigid polycarbonate, wall thickness at sensor position 3 mm
max.
Level display of the sensor holder ±10 mm
Pressure
Measurement range
-200 mm Hg to +800 mm Hg
Resolution
1 mm Hg
Cardioplegia
Pressure measurement range
-200 mm Hg to +800 mm Hg
Resolution
1 mm Hg
Temperature monitor
Display range
0 •C to +50 •C
Timer
Counting range
0 - 999 min 59 sec
Timer (optional)
Counting range
0 - 10 h (up and down)