7 of 28
2019-08-12
Service Record Boogie Fix and Swing
2 General information
2.5 General safety instructions
Check before every use:
•
the firm seat of all pads,
• Frames, attachments and accessories for visible damage, bends, cracks or missing / loose
screws,
•
Wheels / thru axles on firm seat,
•
sufficient tire pressure, tire tread,
• Functioning of the brakes,
•
tight fit of the locking mechanisms (tripod springs, thru axles, eccentric tensioners, etc.),
whether all previously dismantled parts are plugged in again and firmly locked.
There is a risk of injuries (such as bruising) to all rotating, rotating or folding parts, inclu-
ding adjustments, repairs and transport.
All standing aid parts are to be handled properly. Do not throw or drop removable parts!
Before starting the test, repair or adjustment work, clean / disinfect the standing aid and
secure it against tipping over and / or falling down.
Use only original spare parts.
Safety nuts may only be used once. Once loosened safety nuts must be replaced by new
ones.
Only the regular maintenance of all safety-related parts on the standing aid by a qualified
rehab workshop protects against damage and maintains our manufacturer's warranty.
When adjusting the angle of the central center column via telescope and clamping lever,
there must not be any users in the standing aid.
The standing aid may only be used indoors, on level (horizontal) and firm ground and
only by the patient for whom it has been adapted. Outdoor use is not permitted.
Structural changes must not be made without our express permission. They must be exe-
cuted and documented by us or by an authorized rehab technician.