GB
6
Guidance and manufacturer’s declaration – electromagnetic immunity
The UP-X898MD is intended for use in the electromagnetic environment specified below. The customer or the user of
the UP-X898MD should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the UP-X898MD,
including cables, than the recommended separation
distance calculated from the equation appliance to the
frequency of the transmitter.
Recommended separation distance
Conducted RF
3 Vrms
150 kHz to 80 MHz
3 Vrms
d
= 1.2
√
P
IEC 61000-4-6
outside ISM bands
c
6 Vrms
150 kHz to 80 MHz
in ISM bands
c
6 Vrms
Radiated RF
3 V/m
3 V/m
IEC 60601-1-2: 2007
d
= 1.2
√
P
80 MHz to 800 MHz
d
= 2.3
√
P
800 MHz to 2.5 GHz
IEC 60601-1-2: 2014
d
= 2.0
√
P
80 MHz to 2.7 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should
be less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment
marked with following symbol:
IEC 61000-4-3
80 MHz to 2.7 GHz
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Summary of Contents for UP-X898MD
Page 15: ......
Page 27: ......
Page 39: ......
Page 51: ......
Page 63: ......
Page 75: ......
Page 97: ......
Page 109: ......
Page 151: ......
Page 163: ......
Page 175: ......
Page 187: ......
Page 211: ......
Page 223: ......
Page 235: ......
Page 247: ......
Page 289: ......
Page 311: ......
Page 322: ......
Page 323: ......