3
Important EMC notices for use in the medical environments
• The UP-D23MD needs special precautions regarding
EMC and needs to be installed and put into service
according to the EMC information provided in this
instructions for use.
• The portable and mobile RF communications equipment
such as cellular phones can affect the UP-D23MD.
Warning
The use of accessories and cables other than those
specified, with the exception of replacement parts sold by
Sony Corporation, may result in increased emissions or
decreased immunity of the UP-D23MD.
Warning
If the UP-D23MD should be used adjacent to or stacked
with other equipment, it should be observed to verify normal
operation in the configuration in which it will be used.
Guidance and manufacturer’s declaration-electromagnetic emissions
The UP-D23MD is intended for use in the electromagnetic environment specified below.
The customer or the user of the UP-D23MD should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The UP-D23MD uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The UP-D23MD is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
GB