WFM 1125 Option 0A/0B/0C/0D Service Manual
1-3
Specifications
Tables 1–1 through 1–20 list the electrical specifications for the WFM 1125
Digital Television Waveform Monitor Option 0A/0B/0C/0D Option
0A/0B/0C/0D. Performance requirements are generally quantitative and can be
tested by qualified service personnel. Reference information describes useful
operating parameters and typical characteristics and performance.
The performance requirements listed in the electrical specification portion of
these specifications apply over an ambient temperature range of 0
°
C to +40
°
C.
The rated accuracies are valid when the instrument is calibrated at an ambient
temperature range of +20
°
C to +30
°
C, after a warm-up time of 20 minutes.
Electrical Specifications
Table 1-1: Vertical deflection system
Characteristic
Performance requirements
Reference information
Frequency Response
Luminance Channel (Y)
50 kHz to 30 MHz:
≤ ±
0.5 % of the frequency
response at 50 kHz.
37.125 MHz:
≤
-12 dB
Chrominance Channel (P
b
, P
r
)
50 kHz to 15 MHz:
≤ ±
0.7 % of the frequency
response at 50 kHz.
18.5625 MHz:
≤
-12 dB
Low Pass Filter
≤
-10 dB at 15 MHz.
The response at 15 kHz does not vary more
than 0.5 % between FLAT and LPF.
Gain
Frequency response and gain accuracy are
maintained in X5 and X10.
Offset Range
±
350 mV for CH 2 and 3 to CH 1
Cursor Error
≤
0.25 % of full scale
Scale
mV
Resolution of 40 pixel/100 mV (div), Standard
Resolution of 60 pixel/100 mV (div), Full
%
Resolution of 28 pixel/10 % (div), Standard
Resolution of 42 pixel/10 % (div), Full
Summary of Contents for Tektronix WFM 1125
Page 4: ......
Page 6: ......
Page 14: ...Service Safety Summary viii WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Page 20: ...Introduction xiv WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Page 21: ...Specifications ...
Page 22: ......
Page 39: ...Operating Information ...
Page 40: ......
Page 42: ...Preparation for Use 2 2 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Page 56: ...Using the Menu 2 16 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Page 61: ...Theory of Operation ...
Page 62: ......
Page 68: ...Theory of Operation 3 6 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Page 69: ...Performance Verification ...
Page 70: ......
Page 96: ...Performance Verification 4 26 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Page 97: ...Adjustment Procedure ...
Page 98: ......
Page 112: ...Adjustment Procedures 5 14 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Page 113: ...Maintenance ...
Page 114: ......
Page 122: ...Removal and Installation Procedures 6 8 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Page 145: ...Options ...
Page 146: ......
Page 149: ...Electrical Parts List ...
Page 150: ......
Page 152: ...Electrical Parts List 8 2 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Page 153: ...Diagrams ...
Page 154: ......
Page 159: ...Mechanical Parts List ...
Page 160: ......
Page 168: ...10 8 WFM 1125 Option 0A 0B 0C 0D Service Manual Please cut away this area below the lines ...
Page 171: ......
Page 172: ......