Resolution switch setting: (Table 3-1.)
Change the Resolution
To change the resolution or increase the response speed,
follow the procedure below.
1.
Remove the cover (two M3
×
6 screws).
2.
Reset the dip switches on the printed circuit board (Fig.
3-1. S101) by referring to Tables 3-1 and 3-2.
3.
Reinstall the cover and fasten with screws.
4.
Install the unit on the machinery and confirm
operations (counter status, etc.)
•
Responce speed switch setting: (Table 3-2.)
Resolution (
µ
m)
100
50
20
10
Switch setting
S101-1
OFF
ON
S101-2
OFF
ON
OFF
ON
•
Maximum response speed (m/min)
:
(Table 3-3.)
TW (
µ
sec)
Output cycle (kpps)
2.5 (100)
5 (50)
10 (25)
20 (12.5)
Switch setting
S101-4
OFF
ON
S101-3
OFF
ON
OFF
ON
20
240
120
48
24
Tw (
µ
s)
Resolution (
µ
m)
100
50
20
10
2.5
300
300
300
192
10
300
240
96
48
5
300
300
192
96