BCU-100 MM
1-31 (E)
After displaying the above test patterns for 15 seconds once again, the following dialog box is
displayed. Select “Yes” when the test patterns are displayed normally. If not, select “No”.
Upon completion of the resolution check using the first resolution, the following dialog box is displayed.
If you want to continue the resolution check using another resolution, select “Yes”. If you want to
exit the resolution check, select “No”. When “Yes” is selected, the screen returns to the resolution
sub menu. Select the number that you want to check to proceed to the resolution check with another
resolution.
If “No” is selected, the screen returns to the “VIF-40 Board Check” after completion of the internal
processing taking some time.
EDID Read
Reads EDID (Extended Display Identification Data) of the connected monitor.
* The monitor should have been connected earlier beforehand. (The power can be turned off.)
If the readout is performed successfully, the message OK is displayed in the dialog box and the
following messages are saved in the log.
* The following shows an extract of the messages as an example.
==Vender/Product ID/EDID version
EISA manufacturer code
= 4DD9
Product code
= 02D0
Serial number
= 0094EE2B
Week of manufacture
= 19
Year of manufacture
= 2006
EDID Structure version
= 1
EDID Revision
= 3
If readout has failed, the message NG is displayed in the dialog box.
At the same time, the VIF-40 board information is saved in the log file .
* The following shows an extract of the messages as an example.
VIF information
CS1 : exist
Product name
: BC
Board name
: BB
Serial Number
: 01234
FPGA version
: 0104
Board revision
: 0
Board type
: 0
Reg0001
: 82
Reg0002
: c0
Reg0051
: 40
Reg0052
: 00