22
Instruction for use | Supercal 5
4.9 Supercal 5S Installation Considerations (including flowmeter)
Before installation
Check the layout and design data of the installation.
The calculator’s pulse factor and the installation location settings must match the
values indicated on the flow sensor - consult the identification plate!
The permissible temperature operating range of the calculator is 5–55 ºC.
The installation and project provisions must be followed.
The calculator faceplate and identification plates must be legible.
The flow sensor should be mounted BEFORE any control valve to exclude any potential
parasitic influences.
During commissioning, the pipe system must be bled of air. Air in the flow sensor may
affect the measurement.
Use only new and appropriate sealing material.
The watertightness of the different connections should be verified.
Lightning protection has to be guaranteed by the house installation.
Straight sections of piping of 3 DN in flow and return of any flow meter or heat
meter must be provided. For the Supercal 5 S up to DN 40 (qp10) straight sections
of 3 DN are already included in the length of the flow sensor.
4.9.1 Horizontal Mounting Position
Sensor heads MUST be placed to the side +/− 45° of the pipe axis to avoid the influence of
air (top) or dirt (bottom).