General Information: 

Operation 

A programming card is supplied inside the LS-MC, a copy is on page 10.

 

The programming card shows the pre-programmed parameters 

that are held in the memory. 

 

  

Power Up 

On power up, the LS-MC will take a few seconds to initialise, then the display will run through a set-up routine, showing a full set of 
figure eight’s, then the software revision and  finally a full set of zeros whilst the microprocessor identifies the correct return echo and 
confirms this repeats enough to be acceptable. 
 
The instrument is now ready for use, and is set-up using the 3 coloured push buttons. 
 
Refer to the programme menu which is on 

page 10

, and also in the instrument itself on a card tucked between the electronics housing 

and the body. 

 

Display Units 

The LS-MC default value is meters. 

 

  

The display units can be changed using the blue button 

. Simply hold the blue button 

 down for 10 seconds and the units will 

change to the next option in the following sequence : 
Metres to Feet to Inches and back to Metres again. 
 
This is indicated by the position of the decimal point in the displayed value : 
 
 
 
 

 
 
Once the units have changed from default to the next   option, the blue button 

 can be released, in which case the new units selected 

are confirmed, or the blue button 

 can continue to be held down, in which case the next units   option is given after a further three 

seconds.  
 
When the display units have been programmed, please note that you must use the same units when programming in the bottom 
reference, 4 and 20mA points. 
 
The 4-20mA output may be set to operate over all or just a part of the total measuring range. There is no theoretical limit on the 
minimum span of the current output, although in practice a span below 100mm is not recommended. With small spans, be aware that 
a high value of damping will greatly affect the performance of LS-MC.  
 

The 4mA level may be set above or below the 20mA level to suit the monitoring or control equipment. 
  
Note: The LS-MC measures and calculates at all times in metres. The units of display are derived as a last operation using a pre-
programmed conversion factor. 

 

 

 

 

 

Units

Display

M

8

ft.

26.24

in.

314.9

Summary of Contents for LS-MC

Page 1: ...nts the distance between the surface of the liquid being detected and the sensor The LS MC has an integral LCD display 3 push buttons under the lid for security and just 8 parameters to set making it one of the simplest ultrasonic level transmitters to calibrate and use Technical Overview LS MC Ultrasonic Level Transmitter 2 Wire loop powered IP67 Housing 12 to 30Vdc power supply Fast speed of res...

Page 2: ...LCD Beam angle 6 3db Resolution Better than 1mm Accuracy measured range 1 0 meter 5mm 1 0 meter 1 of measured distance Damping 1 to 99 seconds adjustable Process connection 2 BSPT Temperature range 20 to 70 C Operating pressure 0 25 to 3bar max 20 C Material Transducer PVDF Housing Glass filled nylon Dimensions 205 5 x 141mm overall Protection IP67 Conformity EN50081 1 EN50082 2 Country of origin ...

Page 3: ... highest liquid level and 0 88 metres away from walls 4 Do not over tighten the fixing nut only tighten by hand Use the hexagon nut not the housing 5 Keep the LS MC away from high voltage current devices and variable speed drives 6 Shield the LS MC from direct sunlight 7 If used with a liquid where foaming could occur locate the LS MC over an area of clear liquid Failure to comply will result in a...

Page 4: ...for 10 seconds and the units will change to the next option in the following sequence Metres to Feet to Inches and back to Metres again This is indicated by the position of the decimal point in the displayed value Once the units have changed from default to the next option the blue button can be released in which case the new units selected are confirmed or the blue button can continue to be held ...

Page 5: ...can be used as an electronic tape measure With an empty tank or vessel the LS MC will read the distance to the bottom of the tank This can be noted and used when setting b rEF later Press blue button again to scroll the display to show the echo size on a scale of 0 100 It is possible to record a value greater than 100 but do not worry about this With the display in this mode the central cursor wil...

Page 6: ...g menu for the 4mA level The display will show the present value of the 4mA level 2 If this value is correct press the red button then the green button È to scroll on to the next menu option Or Press the blue button to edit this value The first digit of the display will flash and may be changed using green button Once correct press blue button to select the next digit and then green button to edit...

Page 7: ...gits should now be flashing If the new value is incorrect the edit sequence for damping can be started again by pressing the blue button If the new value is correct press the red button to save the damping value programmed in and the display will automatically scroll on to the alarm action menu AL Selecting the output current output on alarm condition The LS MC will signal an alarm condition in th...

Page 8: ...ading is displayed the instrument can be taught to ignore this false echo This routine may be used at any time either during or after set up or should a problem occur later Press blue button to enter the Lrn menu The display will show LrnX where X is 0 1 or 2 and is the number of false target echoes already stored To store a new false echo press the blue button for greater than 5s Note If there ar...

Page 9: ...hing LE with 0000 This means that the LS MC is not receiving a return echo possibly because the liquid surface is poor or beyond the range 8m 26ft of the instrument Re locate the instrument Error messages continued Flashing LE with a level reading This means that the LS MC is no longer receiving satisfactory echoes from the liquid surface This may be because of one of a variety of reasons excessiv...

Page 10: ...vel Default 4 b rEF 20 d Level AL S 4 S 20 Lrn Display Edit Siaplay B o tto m Ref B o tto m Ref B o tto m Ref Display Edit Display 4mA 4mA Level 4mA Level Level Display Edit Display 20mA 20mA Level 20mA Level Level Display Edit Display Damping Damping Damping Display A larm Scro ll Display Current A ctio n Hi Lo Ho ld A ctio n Display Level Flash 4 Display Level fo r 4mA with Level fo r 4mA Displa...