Sontay LS-MC Quick Start Manual Download Page 9

 
 

 

 

 

Page 9 of 10

LS-MC

Date of Issue: 19/09/2013

Issue Number: 5.2

Tel: 

+44 (0) 1732 861200 -

 

E-mail:

 

[email protected] -

 

Web:

 

www.sontay.com

 

© 2012 Sontay Limited. All rights reserved

 

 

Settings: 

Power Failure: 

All parameters are held in E² PROM memory, so in the event of a power failure or disconnection from the power supply the LS-MC will 
remember all of its last parameter values and will resume correct operation once power is restored. 
  

Final Checks: 

 

Check the display is reading correctly. 

 

 

You may wish to check echo size again before re-fitting the enclosure lid. 

 

 

Check that the cover seal is in place in the cover and is good condition. It should not be twisted or kinked in any way. 

 

 

Carefully set the cover on the LS-MC with the text on the cover label at right angles to the conduit entry, and tighten the three 
cover screws equally to seal the instrument. 

 

 

Check that the cable gland is securely tightened and sealing on the cable sheath. 
 

 
 

No display:  

Check power supply. Ensure a minimum of 12v dc at the instrument terminals. Check cable insulation is not preventing contact at the 
terminal block.

 

  

No level reading:  

Check instrument is pulsing. You should be able to hear the transducer “ticking” about once per second. If there is no sound, the 
instrument should be replaced.

 

  

Error messages:  

Flashing “LE” with “0000”. This means that the LS-MC is not receiving a return echo possibly because the liquid surface is poor or 
beyond the range (8m/26ft) of the instrument.

 

Re-locate the instrument.

 

 

Error messages (continued):  

Flashing “LE” with a level reading. This means that the LS-MC is no longer receiving satisfactory echoes from the liquid surface. This may 
be because of one of a variety of reasons, excessive foaming, turbulence or ullage vapours.

 

  
First, check that the instrument transmit face is free from contamination and condensation. The LS-MC will operate with some 
condensation on the face, but excessive condensation may cause operational problems. The vessel should be adequately vented to 
prevent condensation forming. 
  
Check the instrument is still vertically aligned above the liquid surface, then check the echo received size. If the echo size is small (<3), 
re-locate the transducer or modify the vessel so that the LS-MC can operate above a more acceptable area of the liquid surface. 
  
Lost echo (LE) will not be signalled until there has been no return echo for a 10 second period. Within the 10 seconds, the output will 
remain fixed. If the 10 seconds elapses and no satisfactory has been received, the output will drive to the current level selected and the 
display will flash the last valid level reading alternately with “LE”. 
  
If, however, a satisfactory echo is received within the 10 seconds, a new output is established and the LE timer is re-set. 

 

 

 

 

 

 

Fault Finding & Error Messages: 

Summary of Contents for LS-MC

Page 1: ...nts the distance between the surface of the liquid being detected and the sensor The LS MC has an integral LCD display 3 push buttons under the lid for security and just 8 parameters to set making it one of the simplest ultrasonic level transmitters to calibrate and use Technical Overview LS MC Ultrasonic Level Transmitter 2 Wire loop powered IP67 Housing 12 to 30Vdc power supply Fast speed of res...

Page 2: ...LCD Beam angle 6 3db Resolution Better than 1mm Accuracy measured range 1 0 meter 5mm 1 0 meter 1 of measured distance Damping 1 to 99 seconds adjustable Process connection 2 BSPT Temperature range 20 to 70 C Operating pressure 0 25 to 3bar max 20 C Material Transducer PVDF Housing Glass filled nylon Dimensions 205 5 x 141mm overall Protection IP67 Conformity EN50081 1 EN50082 2 Country of origin ...

Page 3: ... highest liquid level and 0 88 metres away from walls 4 Do not over tighten the fixing nut only tighten by hand Use the hexagon nut not the housing 5 Keep the LS MC away from high voltage current devices and variable speed drives 6 Shield the LS MC from direct sunlight 7 If used with a liquid where foaming could occur locate the LS MC over an area of clear liquid Failure to comply will result in a...

Page 4: ...for 10 seconds and the units will change to the next option in the following sequence Metres to Feet to Inches and back to Metres again This is indicated by the position of the decimal point in the displayed value Once the units have changed from default to the next option the blue button can be released in which case the new units selected are confirmed or the blue button can continue to be held ...

Page 5: ...can be used as an electronic tape measure With an empty tank or vessel the LS MC will read the distance to the bottom of the tank This can be noted and used when setting b rEF later Press blue button again to scroll the display to show the echo size on a scale of 0 100 It is possible to record a value greater than 100 but do not worry about this With the display in this mode the central cursor wil...

Page 6: ...g menu for the 4mA level The display will show the present value of the 4mA level 2 If this value is correct press the red button then the green button È to scroll on to the next menu option Or Press the blue button to edit this value The first digit of the display will flash and may be changed using green button Once correct press blue button to select the next digit and then green button to edit...

Page 7: ...gits should now be flashing If the new value is incorrect the edit sequence for damping can be started again by pressing the blue button If the new value is correct press the red button to save the damping value programmed in and the display will automatically scroll on to the alarm action menu AL Selecting the output current output on alarm condition The LS MC will signal an alarm condition in th...

Page 8: ...ading is displayed the instrument can be taught to ignore this false echo This routine may be used at any time either during or after set up or should a problem occur later Press blue button to enter the Lrn menu The display will show LrnX where X is 0 1 or 2 and is the number of false target echoes already stored To store a new false echo press the blue button for greater than 5s Note If there ar...

Page 9: ...hing LE with 0000 This means that the LS MC is not receiving a return echo possibly because the liquid surface is poor or beyond the range 8m 26ft of the instrument Re locate the instrument Error messages continued Flashing LE with a level reading This means that the LS MC is no longer receiving satisfactory echoes from the liquid surface This may be because of one of a variety of reasons excessiv...

Page 10: ...vel Default 4 b rEF 20 d Level AL S 4 S 20 Lrn Display Edit Siaplay B o tto m Ref B o tto m Ref B o tto m Ref Display Edit Display 4mA 4mA Level 4mA Level Level Display Edit Display 20mA 20mA Level 20mA Level Level Display Edit Display Damping Damping Damping Display A larm Scro ll Display Current A ctio n Hi Lo Ho ld A ctio n Display Level Flash 4 Display Level fo r 4mA with Level fo r 4mA Displa...

Reviews: