Operating Manual | 3 Description of the device
SONO-PR 200
25 / 47
Revision: 2.0 | 2022-10-14
Query
Description
Response
LF?\r\n
Low pass filter [LPF] query
LP FILT = x MHz\r\n
ME?\r\n
Mode query
MODE = x \r\n
x = PECH or THRU
PF?\r\n
Pulse repetition frequency [PRF] query
PRF = x KHz\r\n
IM?\r\n
Pulse type query
IMPULSE = x\r\n
x = SPIKE, RECT+, RECT-, RECT
VO?\r\n
Voltage query
Available for “RECT” pulse only.
VOLTAGE = x V\r\n
BU?\r\n
Burst query
Available for
“RECT” pulse only.
BURST = x\r\n
FR?\r\n
Frequency (square pulse) query
Available for “RECT” pulse only.
FREQUENCY = x KHz\r\n
VN?\r\n
Device identification query
SONO-PR 200 VERSION x\r\n
x = firmware version number
Examples
The following table contains examples of queries and responses:
Query
Response
PF=?\r\n
PRF = 10.00 KHz\r\n
PR=?\r\n
ER\r\n