Operating Manual | 3 Description of the device
SONO-PR 200
24 / 47
Revision: 2.0 | 2022-10-14
Command
Response
EY=32\r\n
OK\r\n
EY=12\r\n
ER\r\n
3.8.3
Queries
Description
The RS-232 interface may be used to send queries for determining the parameter values
currently set in the SONO-PR 200.
After sending a query, the SONO-PR 200 will return one of the following responses:
Query
Response
Valid
Queried parameter including value
Example: PRF = 10.00 KHz\r\n
Invalid
ER\r\n
For sending a query, the respective parameter must be used in the following syntax:
PARAMETER=?\r\n
Available queries and responses
The following table contains a list of all available queries and the respective responses:
Query
Description
Response
DG?\r\n
Damping query
DAMPING = x Ohm\r\n
EY?\r\n
Energy query
Available for “
SPIKE
” pulse only.
ENERGY = x uJ\r\n
GN?\r\n
Gain query
GAIN = x dB\r\n
HF?\r\n
High pass filter [HPF] query
for HPF = 100 kHz:
HP FILT = 100 KHz\r\n
for HPF > 100 kHz:
HP FILT = x MHz\r\n
IA?\r\n
Input attenuation query
ATTENUATORS = x dB\r\n
OA?\r\n
Output attenuation query
ATTENUATORS = x dB\r\n