Operating Manual | 3 Description of the device
SONO-PR 200
18 / 47
Revision: 2.0 | 2022-10-14
No. Type
Description/function
Trigger
Switching the signal triggering mode:
INTERN (internal trigger
–
PRF)
EXTERN (external trigger)
Frequency (Width)
Setting the pulse width:
Range: 63 …
1000 ns
Available for “RECT” pulse only.
2
Brightness
Adjusting screen brightness.
Opens the “Brightness” screen.
3
Serial number
Shows the device's serial number.
4
Device configuration
Shows the device name and configuration (Spike or Combi).
Automatic constraint of pulse repetition frequency (PRF)
When using the “RECT” pulse type, it will not be possible to simultaneously set
maximum values for
“Voltage”, “Damping”, “Burst”, “Width” and “PRF”. Depending on the
particular settings, the software will automatically constraint the PRF. The maximum
permissible PRF for the current parameter settings is displayed the “Settings” screen.