SONOFLOW IL.52 V2.0
Operating Manual
12
Revision: 1.0; Date: 2018-10-26
5
Output Configuration and Settings
The settings can be adapted for special measuring tasks using the SONOFLOW Monitor
(optional accessory).
5.1 Current output for flow rate
The default settings for current output are:
Parameter
Default setting
Further options
(via SONOFLOW Monitor)
Configuration
Flow
Off |
Test current (
0 … 22 mA)
Range
4 … 20 mA
0 … 20 mA
Limit
> 20 mA
≤ 20 mA
Flow min.
0 ml/min (4 mA)
0 … 3 000 ml/min
Flow max.
3 000 ml/min (20 mA)
0 … 3 000 ml/min
Measuring Error
2 mA
0 … 22 mA
Device Fault
0 mA
0 … 22 mA
Table 3: Default settings for current output
5.2 Frequency output for flow rate
Parameter
Default setting
Further options
(via SONOFLOW Monitor)
Configuration
Flow
Off |
Test frequency (
0 … 20 kHz)
Flow min.
0.00 ml/min (0 kHz)
0 … 3 000 ml/min
Flow max.
3000 ml/min (20 kHz)
0 … 3 000 ml/min
Max. frequency
20 kHz
1 … 20 kHz
Measuring Error
0 kHz
0 … 20 kHz
Device Fault
0 kHz
0 … 20 kHz
Table 4: Default settings for frequency output