Cleaning, disinfecting, transporting, storing, and disposing of the transducer
4-9
Do not stack items on top of the case.
Do not exceed the shipping temperature range: -35° C (-31° F) to +65° C (149° F).
Do not open the case until it reaches its final destination.
After arrival, the transducer must be cleaned and disinfected using the procedures detailed in
on page 4-1 before it can be used.
Storing the transducer
To store the transducer
1
Clean and disinfect the TEExi transducer. See
2
Store the transducer so that it hangs freely and vertically, and observe the following precautions:
Store the transducer away from any contaminated transducers.
Store the transducer in an environment that is safe and has good airflow. Do not store the transducer
in closed containers or where condensation may occur.
Use a tip cover when storing the transducer to prevent damage to the scan head. The tip cover
encloses and protects the scan head from mechanical strain and impact during storage. Keep the tip
cover on until preparing the transducer for use.
Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between
0° C (32° F) and +45° C (113° F).
If using a wall-mounted rack for storage, ensure that:
It is securely mounted.
The storage slots do not mar the transducer or the endoscopic shaft.
The rack is sized and positioned to prevent the transducer from inadvertently falling.
Make sure the connector is supported and secure.
Disposing of the transducer
Caution
The tip cover is a single-use device. Do not reuse tip covers. Doing so can
result in contamination of, or damage to, the transducer.
WARNING
Do not destroy the transducer by incinerating or burning it. Return the
transducer to FUJIFILM SonoSite or your local representative for disposal.