Examination
3-3
In addition to the high level disinfection, a protective sheath may provide even greater protection against
contamination of the transducer. Contact CIVCO for protective sheaths and applicators for protective
sheaths.
Bite guard
Each TEExi transducer from FUJIFILM SonoSite is delivered with three bite guards. Use of a bite guard is
mandatory. If you need help ordering more bite guards, contact CIVCO Medical Solutions.
Re-use, cleaning, and sterilization of the bite guards should be done according to instructions provided by
the manufacturer of the bite guard.
Figure 3-1
Bite guard: Side view (left) and front view (right)
Sterile sheath
Use a sterile sheath whenever examining a patient that poses an isolation risk.
There are various sterile sheaths available to eliminate direct contact between the patient and the
endoscope. Follow the user instructions for a particular sheath when applying and removing the sheath from
the TEExi transducer. Contact CIVCO to order sterile sheaths and applicators for sterile sheaths.
To provide suitable acoustic coupling within the sheath, FUJIFILM SonoSite recommends using a sterile gel.
Caution
To avoid damaging the transducer, use a bite guard during all TEExi examinations.
Biting the endoscope may cause severe, permanent damage to the transducer,
making it unsafe for future patient use.
Damage to the transducer from failure to use a bite guard voids the transducer
warranty.
Caution
To avoid damaging the TEExi transducer, ensure that its tip is straight during
application and removal of the sheath. During removal of the sheath, be careful
not to use excessive force on the transducer tip; otherwise permanent damage to
the TEExi transducer may occur.