2-6
Getting Started
TEExi transducer controls
The endoscope is designed for one-hand operation of the deflection and scan plane controls.
shows the user holding the endoscope handle in the left hand. Thumb and first and second fingers operate
the deflection and scan plane controls.
Check the mechanical operation and physical integrity of the transducer after taking it out of the box and
before each exam.
Figure 2-4
Transducer in left hand
To visually and tactilely inspect the transducer
You should inspect the TEExi transducer visually and tactilely after taking it out of the box and before
disinfecting.
1
Visually examine and feel the entire surface of the flexible shaft and deflection section with the transducer
in both straight and deflected position.
2
Articulation section
8
Transducer connector
3
Transducer tip with scan head
9
Scan plane control buttons
4
Deflection brake
10
Attachment ring
5
Deflection control wheels
11
Handle
6
Neutral marker
WARNING
To avoid injury to the patient:
Do not use the TEExi transducer if any irregularity, substandard function or
unsafe condition is observed or suspected.
Do not use the TEExi transducer if any metallic protrusions, holes, rough spots,
cracks, or dents are found.