background image

Schematic Diagram, Docking Stand PCB

SiteStand Service Manual

15

A.5

Schematic Diagram, Docking Stand PCB

Summary of Contents for SiteStand

Page 1: ...SonoSite SiteStand Mobile Docking Station Service Manual ...

Page 2: ... ii P01117 01 Rev B 10 00 Copyright 2000 by SonoSite Inc All rights reserved Printed in the USA ...

Page 3: ...Non SonoSite product names may be trademarks or registered trademarks of their respective owners SonoSite products may be covered by one or more of the following U S patents 4214269 4387365 4387597 4432371 4435614 4454884 4534357 4542653 4552607 4581636 4603702 4607642 4644795 4670339 4773140 4817618 4883059 4887306 5016641 5095910 5099847 5123415 5158088 5197477 5215094 5226420 5233994 5255682 52...

Page 4: ... iv ...

Page 5: ...troduction 7 3 1 1 Contacting SonoSite Technical Support 7 3 2 Troubleshooting the SiteStand 7 3 3 Disassembling the SiteStand 8 3 3 1 Required Tools 9 3 3 2 Replacing the Sleeve Assembly and the Docking PCBA 9 3 3 3 Replacing the Power Supply 9 3 3 4 Replacing the Foot Pedal 9 3 3 5 Repairing Replacing the Front Panel Connectors 10 3 3 6 Replacing the SiteStand Fuse 10 3 4 SiteStand Performance T...

Page 6: ... 2 Replacement Parts List 13 A 3 Ordering Replacement Parts 13 A 4 SiteStand Wiring Diagram 14 A 5 Schematic Diagram Docking Stand PCB 15 A 6 SiteStand Assembly 16 A 7 SiteStand Sleeve Assembly 17 APPENDIX B Service Event Report 21 ...

Page 7: ...ilities for the system The SiteStand has storage for two transducers and a tray for a recommended black and white printer It provides the following connections three video ports an RS 232C port a printer control port and two AC mains IEC power receptacles You can also tilt the system and adjust the height of the system when it is in the docking station 1 2 Setup and Operation For setup and operati...

Page 8: ...Chapter 1 Introduction 2 SiteStand Service Manual Figure 1 1 SiteStand with Display SiteStand display SonoSite system SiteStand Printer tray Height adjustment pedal Locking levers Sleeve ...

Page 9: ... and 100 120 220 240 VAC 50 60 Hz 2 0 A 2x AC power adapter input 100 120 220 240 VAC 50 60 Hz 1 0 0 50 A AC power adapter output 16Vdc 2 8 A 1 3 2 SiteStand Display 1 3 2 1 Display Type TFT LCD 640 x 480 pixels 1 3 2 2 Electrical SiteStand flat panel display input 12VDC 2 75 A AC power adapter input 100 120 220 240 VAC 50 60 Hz 0 85 A maximum AC power adapter output 12VDC 2 75 A ...

Page 10: ...Chapter 1 Introduction 4 SiteStand Service Manual ...

Page 11: ...ock and fire hazard inspect the AC power adapter cord and plug on a regular basis Ensure they are not damaged Connection of peripherals not recommended by SonoSite could result in electrical shock Avoid electrical shock hazards by using peripherals and accessory cables recommended by SonoSite To avoid the risk of electrical shock use commercial grade peripherals recommended by SonoSite on battery ...

Page 12: ...ut the flat panel display please refer to the Operating Instructions for the Series DV 3000 Colour LCD Rugged Monitor 2 1 1 Equipment Protection To protect your ultrasound system transducer and accessories follow these precautions CAUTIONS Excessive bending or twisting of cables can cause a failure or intermittent operation Improper cleaning or disinfecting of any part of the system can cause perm...

Page 13: ...chnical support or to order spare parts do one of the following For U S customers call 1 877 657 8118 For international customers call 425 951 1330 Connect to SonoSite on the World Wide Web at www sonosite com Select Products then choose Technical Support E mail service sonosite com 3 2 Troubleshooting the SiteStand The intent of this section is to help isolate potential problems associated with t...

Page 14: ... refer to the SonoSite 180 Service Manual for troubleshooting There is no video from the SiteStand video out connection The SiteStand docking PCBA has failed The docking connections or the sleeve assembly have been broken See Replacing the Sleeve Assembly and the Docking PCBA on page 9 There is no AC power from the SiteStand power connection The SiteStand fuse has failed See Replacing the SiteStan...

Page 15: ...nch with the base facing you 2 Use an Allen wrench to remove the four 5 32 inch Allen screws holding the base 3 Remove the base and set it aside 4 Use an Allen wrench to remove to remove the six 5 32 inch Allen screws and the 11 16 inch nut on the gas spring securing the subplate 5 Remove the subplate 6 The power supply is visible in the forward section of the SiteStand and is secured by Velcro Pu...

Page 16: ...n place 4 Use an Allen wrench to remove the eight 5 32 inch Allen screws from the top subplate 5 Left the blade cap and subplate to the top of the shaft and secure it in place 6 Gently pull the connector assembly out the top of the blade assembly 7 The connector can now be repaired or replaced 8 Reassemble the SiteStand by performing steps 1 through 8 in this section in reverse order 3 3 6 Replaci...

Page 17: ...te 180 or SonoHeart without a battery into the docking station Apply AC power 120V 60Hz to SiteStand AC in receptacle 3 4 2 AC Power Verification Use an AC Tester to verify both IEC output receptacles are wired correctly 3 4 3 Docking Video Turn on the system Connect a Test Monitor to each of the video outputs on the docking stand Verify the same video pattern exists on the Test Monitor as is pres...

Page 18: ...Chapter 3 Servicing the SiteStand 12 SiteStand Service Manual 3 Repeat steps 1 and 2 above at least one more time to verify that the status line is connected ...

Page 19: ...tic and a wiring diagram are provided for the SiteStand and the docking stand PCB A 2 Replacement Parts List The following tables contain all the replaceable parts for the SiteStand All quantities are one unless otherwise noted A 3 Ordering Replacement Parts To order parts contact SonoSite technical support at 1 877 657 8118 U S or 425 951 1330 international ...

Page 20: ...Input 100 240V 50 60Hz Power Adapter Output 16VDC CSA UL approved to IEC601 Based Standards Docking Stand PCB 16VDC 16VDC RTN 1 V 2 V 3 Video 4 GND 5 Print Control 6 Print Status 7 GND 8 Rx 9 Tx 10 GND 11 Rts 12 Cts 13 Sense 14 Spare NOTE Docking Stand is disabled with no voltages at 14 output contacts until SonoSite 180 is placed in Stand connecting V to Ground SonoSite 180 Printer Control Status...

Page 21: ...Schematic Diagram Docking Stand PCB SiteStand Service Manual 15 A 5 Schematic Diagram Docking Stand PCB ...

Page 22: ...Appendix A Schematic and Parts List 16 SiteStand Service Manual A 6 SiteStand Assembly ...

Page 23: ...SiteStand Sleeve Assembly SiteStand Service Manual 17 A 7 SiteStand Sleeve Assembly ...

Page 24: ...4 P01562 washer lock 1 4 inl stl 12 1 P00987 base caster 13 4 P01532 washer lock 1 2 split sstl 14 2 P00693 caster 15 2 P00694 caster nonlocking 16 1 P00994 pedal actuator 17 1 P00995 tube vertical 18 1 P00986 housing bearing 19 2 P00984 ring bearing 20 2 P00983 ring retainer spring 21 1 P00996 plug float 22 1 P00678 gas spring 23 1 P00991 collar tilt 24 2 P00980 bushing flanged 1 4x 377x 25lg tef...

Page 25: ...acket 39 1 P01629 nut m10 x 1 25 x 7 78 40 1 P00409 assembly PCB docking stand 41 1 P01595 assembly bracket wiring harness dock stand 42 1 P00590 cover connector 43 6 P01566 screw phpls 6 32x 50lg 44 1 P01318 shoulder bolt 1 1 4 inch 45 1 P01319 locknut nylon insert 10 24 1 P00790 fuse 6 3 amp 5 x 20 mm 1 P01321 foot pedal repair kit P00762 printer control cable P00764 printer AC power cable P0076...

Page 26: ...ve 5 1 P00663 e ring stand 6 1 P00650 cover latch mechanism 7 2 P00766 spring torsion dock 8 1 P00665 probe insert right 9 1 P00666 probe insert left 10 15 P00787 screw k35x1 57x8 type 1 recess panhead tf 11 1 P00667 screw k35x20 cross recess panhead tf 12 1 P00802 washer shoulder nylon length 13 32 13 2 P00800 washer flat 12 ss finishing 14 1 P00669 spring compression lid 15 6 P00594 pad poron 09...

Page 27: ...de approved warranty services for SonoSite products For all repairs completed complete the form and return a copy of it to the following address SonoSite Inc Technical Support 21919 30th Drive SE Bothell Washington 98021 3904 telephone 1 877 657 8118 U S 425 951 1330 international facsimile 425 951 1416 e mail service sonosite com website www sonosite com Select Products then Technical Support ...

Page 28: ...fiers Event Description Diagnosis Service Performed Performed By Date Actions Parts Removed Part Name Part Number Serial Number Lot Number Rev Replaced By Parts Installed Part Name Part Number Serial Number Lot Number Rev Replaced By Tests Performed attach test data Test Test Performed By Performed By Result Pass Fail Result Pass Fail Attach additional sheets as required Page ____ of ____ F00019 R...

Page 29: ... sleeve assembly 20 product failures reporting 21 R replacement parts list 13 ordering 13 S safety 5 electrical 5 equipment protection 6 schematic 13 service event report 21 form 22 SiteStand assembly drawing 16 description 1 figure 2 performance tests 11 replacing the fuse 10 servicing 7 setup and operation 1 troubleshooting 7 SiteStand sleeve assembly drawing 17 specifications SiteStand 3 SiteSt...

Reviews: