SONOPAN LMC-10 Instruction Manual Download Page 6

 

 

 

 

- 4 - 

 

Fig. 3. Typical spectral sensitivity of the L 200 probe. 

 

1.3.2.  Directional characteristics 

The unique optical  system  of  the  measurement  probe  ensures  an  identical  assessment 
of every point of the tested surface. The special attention should be paid to the fact that 
the countersink angle for the cone of rays reaching the detector from every point of the 
tested  object  is  identical  with  the  measurement  field  angle,  which  is  unachievable  for 
standard luminance meters and can become the source of serious measurement errors 
for  them.  Such  shape  of  the  beam  of  rays  is  the  main  attribute  for  the  LMC-10 
measurement  probe  and  its  enormous  advantage  over  other  meters  available  on  the 
market.  Therefore,  the  influence  of  luminance  irregularity  of  the  monitor  matrix  for 
various directions of observations is negligible in the case of the LMC-10 meter.  

 

Fig. 4. Directional response of the LMC 10 probe. 

 

1.4.  Temperature compensation 

The change in the temperature of the probe influences the measurement result. Due to 
the temperature change, the other parameters also change and they are as follows :  

 

detector dark current 

 

 

- shift of zero indication,  

 

offset voltage of the measuring circuit  - shift of zero indication,  

 

responsivity of a detector   

 

- change of measured quantity value,  

Summary of Contents for LMC-10

Page 1: ...vember 2018 CONTACT LUMINANCE METER LMC 10 INSTRUCTION MANUAL PPUH SONOPAN sp z o o ul Cio kowskiego 2 2 15 950 Bia ystok Poland phone fax 48 85 742 36 62 sonopan sonopan com pl http www sonopan com p...

Page 2: ...f zero indication 5 1 4 2 Temperature compensation of the result 5 2 THE LMC 10 METER OPERATION 5 2 1 Preparations 5 2 2 Measurement 5 3 BASIC FUNCTIONS OF CONTROL UNIT 6 4 EXTENDED FUNCTIONS OF CONTR...

Page 3: ...sen measurement procedure e g to test the reproduction of the grayscale of monitors according to DICOM GSDF1 function The function of the key can be dependent on the running application of the control...

Page 4: ...surement probe P 200 measurement module LMC 10 measurement probe RF 200C interface RF200C measurement module Each of the above mentioned components of the measurement system is identifiable according...

Page 5: ...s realised with the use of the set of glass absorption filters Such a solution guarantees high long term and temperature stability especially in comparison with the cheaper realisations in which inter...

Page 6: ...em Such shape of the beam of rays is the main attribute for the LMC 10 measurement probe and its enormous advantage over other meters available on the market Therefore the influence of luminance irreg...

Page 7: ...use of the high class elements in the design of the meter the LMC 10 probe is equipped with the automatic compensation system of the influence of the environmental temperature on the measurement valu...

Page 8: ...The control devices are equipped with applications or functions that extend the range of the probe usage All the applications have the possibility of archiving of the measurements results to an exter...

Page 9: ...trol units 4 4 Application Advanced 4 4 1 Relative measurement The control units enable to conduct and archive the measurements whose results are expressed as a percentage of any reference It is possi...

Page 10: ...a vertical position The probe cap should be removed only during the measurement The probe should be removed and put on while turning it right clockwise which prevents it from being untwisted by accide...

Page 11: ...c compatibility The conformance of the above mentioned requirements is confirmed by CE mark This product cannot be thrown away with household waste Deposit the product in an authorized electrical and...

Reviews: