- 1 - 

1. CHARACTERISTICS  

The  SONOPAN LMC-10  contact  luminance  meter  is  designed  for  the  luminance 
measurement  of  the  self-luminescent  surfaces  such  as  monitor  matrices,  TV  sets, 
negatoscopes,  reading  panels,  etc.  The  meter  ensures  the  accurate  measurement, 
regardless  of  the  spatial  distribution  of  the  measured  luminance  and  the  backlight 
system of the surface being tested. 

It is indispensable for the control of the medical 

imaging  monitors  and  negatoscopes.

  Its  unique  optical  and  measurement  systems  in 

combination with the advanced algorithms of the measured signal processing guarantee 
high accuracy and repeatability of measurements.  
Owing  to  its  technical  parameters,  the  LMC-10  meter  is  classified  as  a 

class  A 

meter 

according to DIN 5032-7 and TC-2.2 CIE.   
The  meter  consists  of  the  measurement  probe  cooperating  directly  with  the 
SONOPAN P-200 control unit and with the SONOPAN RF-200C and the USB with a PC or 
a tablet. The measurement probe is equipped with the multifunctional key that is used 
to trigger a measurement, save results and operate the chosen measurement procedure 
e.g.  to  test  the  reproduction  of  the  grayscale  of  monitors  according  to  DICOM  GSDF

 1)

 

function.  The  function  of  the  key  can  be  dependent  on  the  running  application  of  the 
control unit.  

 

 

1.1.  Equipment  

1.1.1.  Basic accessories   

Components of set 

Set with  

P-200 control unit 

Set with PC 

LMC-10 measurement probe 

Stabilizing cap Ø50mm 

P-200 control unit 

– 

RF-200C interface 

– 

USB cable 

– 

RF200C software 

– 

Instruction manual 

CE Declaration of Conformity 

Warranty card 

Carrying case 

 

1.1.2.  Additional accessories 

Tablet 10" (concerning the set with a PC).  
 

1.2.  CONFIGURATION  

The  SONOPAN  P-200  control  unit  or  the  RF200C  software  can  be  used  to  control  the 
LMC-10 contact luminance meter and read the measured values. In the former case, the 
probe is connected directly to the control panel while in the latter case, it is attached to 
the USB 2.0 port with the use of the RF-200C interface. As far as PCs are concerned, it is 
required  to  use  Windows XP  or  higher.  As  for  tablets,  they  should  be  equipped  with 
Windows 10. The RF200C does not work yet with such systems as Windows CE, Android 
or iOS.  
                                                     

1)

  DICOM GSDF: Grayscale Standard Function according to PS 3.14-2011 Digital Imaging and 

Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function. 

Summary of Contents for LMC-10

Page 1: ...vember 2018 CONTACT LUMINANCE METER LMC 10 INSTRUCTION MANUAL PPUH SONOPAN sp z o o ul Cio kowskiego 2 2 15 950 Bia ystok Poland phone fax 48 85 742 36 62 sonopan sonopan com pl http www sonopan com p...

Page 2: ...f zero indication 5 1 4 2 Temperature compensation of the result 5 2 THE LMC 10 METER OPERATION 5 2 1 Preparations 5 2 2 Measurement 5 3 BASIC FUNCTIONS OF CONTROL UNIT 6 4 EXTENDED FUNCTIONS OF CONTR...

Page 3: ...sen measurement procedure e g to test the reproduction of the grayscale of monitors according to DICOM GSDF1 function The function of the key can be dependent on the running application of the control...

Page 4: ...surement probe P 200 measurement module LMC 10 measurement probe RF 200C interface RF200C measurement module Each of the above mentioned components of the measurement system is identifiable according...

Page 5: ...s realised with the use of the set of glass absorption filters Such a solution guarantees high long term and temperature stability especially in comparison with the cheaper realisations in which inter...

Page 6: ...em Such shape of the beam of rays is the main attribute for the LMC 10 measurement probe and its enormous advantage over other meters available on the market Therefore the influence of luminance irreg...

Page 7: ...use of the high class elements in the design of the meter the LMC 10 probe is equipped with the automatic compensation system of the influence of the environmental temperature on the measurement valu...

Page 8: ...The control devices are equipped with applications or functions that extend the range of the probe usage All the applications have the possibility of archiving of the measurements results to an exter...

Page 9: ...trol units 4 4 Application Advanced 4 4 1 Relative measurement The control units enable to conduct and archive the measurements whose results are expressed as a percentage of any reference It is possi...

Page 10: ...a vertical position The probe cap should be removed only during the measurement The probe should be removed and put on while turning it right clockwise which prevents it from being untwisted by accide...

Page 11: ...c compatibility The conformance of the above mentioned requirements is confirmed by CE mark This product cannot be thrown away with household waste Deposit the product in an authorized electrical and...

Reviews: