Instructions for Assembly and Operating
SONIX 30D
12
Password chan
F12 Starting the
backlight
F3 Change of the
current output
properties
F2 Setting the
thresholds
F6 Change of
display precision
F9 Calibration of
the current
output
F15
Amplification
Switching on/off the illumination of LCD display.
Programming of the lower and upper flow threshold in m
3
/h or t/h.
Programming from 0 to 3 decimals for temporary flow and the totalized
volume.
Forcing the 0, 4, 10, 15 or 20mA current in order to achieve a calibration of the
device connected to the current output of the flowmeter.
Six 3-digit numbers are visible on the screen. Each digit represents the
amplification of the signal from the probe.The smaller the amplification the
stronger signal from the probe. The correct values should be in range
100...500. When the amplification is greater then 750 it means that the signal
is too small. This may be due to sediment on the probe's surface or damage
of the probe.
The function allows dosing the liquid. The temporary liquid counter is set to
zero and initiated with the left push button and stopped with the right one.
The counter may also be operated by the RS485 series link (see Protocol of
series transmission).
F8 Temporary
volume
Change of output current: 0-20mA or 4-20mA
Programming the characteristics of the current output:
1 – the output works symmetrically for the negative and positive flows
0 – the output work only for the positive flow
bottom line:
1 – the current 4..0mA responds to flow 0..-0,25Q
p
(only for 4..20mA range)
F4 Parameters of
the RS485 output
Programming parity control: Odd, Even, None, address of the device: 01
H
...FF
H
and transmission speed: 1200B/s...9600B/s.
F5 Time constant
Changing 0...9 responds to app. 0,5...30s time constant.
Time constant is separated from averaging mode (F14).
This funkcion is unaccesible after verification.
F10 Calibration
of the pulse
output
Forcing the pulses in order to achieve a calibration of the device connected to
the pulse output of the flowmeter.
The number of pulses is in accordance with the programmed pulse unit (F13)
and is shown as a percentage of the range flow Q
p
.
This funkcion is unaccesible after verification.
F7 Access code
change
Allows user to change 4 digits access code to user funkcions.
Default access code: 1111
F11 Language
selection
Menu language selection. Available are: Polish, English, Swedish
F13 Pulse unit
change
Pulse output unit change in m
3
. Programmed unit must not be less than the unit
displayed in bottom line. Pulse unit change is not available after verification.
F14 Averaging
Selection of the averaging method. Several methods are avaiable:
- arithmetic mean
- median
- extended median
- SONIX filter
This funkcion is unaccesible after verification.