background image

 

 

 

Page 

14

 of 

22

 

 

Time 

- : --: --  

Pulse 

-- --  

 

Ampl 40 %

 

 

NOTE: To clear an incorrect entry 

press the CLEAR key. 

7.3. 

OPERATION 

Press the 

I

 

key. The screen will display the power rating of the Ultrasonic Processor and the 

following control parameters. 

 

 

 

 
 
AMPLITUDE:

 Desired amplitude must be set in order for the Ultrasonic Processor to be 

operational.  AMPL displays the amplitude selected (e.g. 40%). To set the amplitude at 40% 
when the ultrasonics is off, press the AMPL key and the numeric keys for a 40% reading on the 
screen, and then press the ENTER/REVIEW key.   
 
Note: The minimum amplitude setting is 20%. 

 

The screen will display:

 

 

 

 

 

 

 

1.  Immerse the probe approximately halfway into the sample.  If the probe is immersed to an 

insufficient depth, air will be injected into the sample, causing the sample to foam.  Also 
ensure that the probe tip is not touching the wall of the sample vessel.  
 

2.  The Ultrasonic Processor is now ready for continuous operation.  To energize the 

ultrasonics, press the push button on the converter. To de-energize the ultrasonics, release 
the push button on the converter.   

 

3.  To increase or decrease the amplitude in small increments when the ultrasonics is on, 

depress the AMPL to display Amplitude Setting on the screen, then depress the 

▲ or ▼ 

key, as required.  Since the amplitude required is application dependent and subject to the 
volume and composition of the sample, it is recommended that the amplitude be selected 
through experimentation by increasing or decreasing the level of intensity as needed to 
properly process the sample to achieve desired results.  
 

 
 
 
REVIEW: 

 

The REVIEW function provides a “window” on the process by displaying various 

operating parameters without process interruption.  Pressing the ENTER/REVIEW key 
repeatedly during processing will consecutively display the following information.

  

 

Summary of Contents for VCX130 PB

Page 1: ...Sonics Materials Inc 53 Church Hill Road Newtown CT 06470 USA Phone 203 270 4600 info sonics com Rev 4 21 www sonics com ULTRASONIC PROCESSOR Part No VCX130 PB OPERATION MANUAL...

Page 2: ...CTIONS OF KEYS CONTROLS CONNECTORS 10 5 4 CONVERTER CLAMPING 11 6 PREPARATION FOR USE 12 7 OPERATING INSTRUCTIONS 13 7 1 CAUTION 13 7 2 SETUP 13 7 3 OPERATION 14 7 4 PUSH BUTTON CONTROL 15 8 MAINTENAN...

Page 3: ...rranty of either merchantability or fitness for a particular purpose No person or company is authorized to change modify or amend the terms of this warranty in any manner or fashion whatsoever Under n...

Page 4: ...ic Processor NEVER immerse the converter in liquids of any kind or let condensed moisture or liquid drip into the converter NEVER grasp an activated probe or touch the tip of a vibrating probe It can...

Page 5: ...ps Slow Blow 1 6 Amps Slow Blow Weight 10 lbs 4 5 Kg Dimensions 9 75 W x13 L x 4 75 H 248mm x 331mm x 121mm Output Voltage 380 VRMS max Output Frequency 20 KHz Converter Weight 0 75 lbs 0 34Kg Dimensi...

Page 6: ...t configured to match the wall receptacle a properly grounded universal AC socket adapter must be added Important Universal adapters do not convert voltage or frequency Manufacturer is not responsible...

Page 7: ...rocessing The frequently asked questions FAQ section has more information on this subject Please consult with a product specialist for assistance with selecting a probe for your application Relationsh...

Page 8: ...the sample can readily be determined by visual observation the amount of power required cannot be predetermined A sensing network continuously monitors the output requirements and automatically adjust...

Page 9: ...Page 9 of 22 5 DESCRIPTIONS OF COMPONENTS FUNCTIONS OF CONTROLS 5 1 VCX130 PB FRONT PANEL 5 2 VCX130 PB REAR PANEL Converter Push Button...

Page 10: ...of vibration at the probe tip key Used with the AMPL key when the unit is on stand by to set the amplitude of vibration at the probe tip Also used to increase or decrease the amplitude in small increm...

Page 11: ...of 22 5 4 CONVERTER CLAMPING Improper clamping can damage the system and void the warranty Proper Clamping Improper Clamping Using a stand and clamp will ensure a proper fit Clamp and Stand Part no 83...

Page 12: ...operation proceed as follows 1 Ensure that the power cord is not connected to the electrical outlet 2 Open the fuse holder cover using a small screwdriver 3 Pull out the red fuse holder from its hous...

Page 13: ...to prevent overheating 7 2 SETUP 1 Plug the electrical line cord into the electrical outlet If the unit is already on as indicated if the display lights up depress the O key 2 Probes must be properly...

Page 14: ...be tip is not touching the wall of the sample vessel 2 The Ultrasonic Processor is now ready for continuous operation To energize the ultrasonics press the push button on the converter To de energize...

Page 15: ...Note The amount of energy displayed will be only for one cycle Initiating a new cycle will reset the display to zero e g 20 watts 0000000 Joules d Elapsed time since processing was initiated e g Elap...

Page 16: ...gged 8 1 PROBE MAINTENANCE Ultrasonic processors create high intensity vibration which puts stress on the converter and probe assembly The sides and end of the probe must never be allowed to come in c...

Page 17: ...emove any debris on the threading of the connecting stud Allow the alcohol to dry completely 3 Clean threading in converter Use alcohol and a cotton swab to remove any debris on the threading Do not a...

Page 18: ...depressing the O key and disconnect the line cord from the electrical outlet 2 Open the fuse holder cover using a small screwdriver and pull out the red fuse holder from its housing 3 Replace the fuse...

Page 19: ...against possible damage in shipment The Ultrasonic Processor should be sent to the address below with all transportation charges prepaid and return of shipment indicated RMA Sonics Materials Inc 53 C...

Page 20: ...o radiological biological or chemical agents and has been decontaminated rendering it safer for handling use or disposal Authorization By accepting authorization to return the equipment listed above t...

Page 21: ...the liquid is pushed down and away in all directions If the vessel is too wide it will not mix effectively and some sample will remain untreated at the periphery The probe should never touch the side...

Page 22: ...power is most critical when trying to reproduce ultrasonic processing results Power has a variable relationship with amplitude intensity For example sonicating water requires less wattage when compare...

Reviews: