Sonics 830-00905 Operation Manual Download Page 4

 

 

 

R

EV

.

  

12-21

 

 

SAFETY

 

PRECAUTIONS

 

AND

 

SYMBOLS  _________________________  

 

   

Read the MSDS for the coolant used and follow 

all

 safety precautions 

listed in the MSDS prior to removing coolant tubes or opening the fill 
cap as this could result in contact with the coolant inside. 

 
 

Caution!  Risk of electric shock.  Disconnect the power cord prior to 
servicing.  This includes changing a fuse or opening the cover for any 
reason.  

 

 

CAUTION 

 

Never disassemble the chiller as irreparable damage may occur. 

 

Any attempt to open or repair the unit will void the warranty 

 

Never store the chiller over 70 °C. 

 

Never operate the chiller in ambient temperatures of 40 ºC or greater unless the unit has been 
customized for high ambient operation. 

 

Never operate the chiller within 5 ºC of the coolant’s freezing point. 

 

Only water is recommended when using a #830-00905 with an ultrasonic processor and cup 
horn/microplate horn accessory.  When using a #830-00906 always use only proper coolants as 
specified in manual.  Koolance LIQ-702CL-B (27% propylene glycol and water) is recommended. 

 

Never ship the chiller with coolant inside the liquid cold plate as freezing temperatures may be 
encountered which would damage the unit.  Always pump all coolant out of the chiller prior to 
shipping.  

 

Symbols Used in this Manual 

 

 

The red CAUTION equilateral triangle symbol appears throughout 
the manual.  Please follow the important instructions accompanying 
this symbol to avoid significant damage to the chiller.       

 

 

The red WARNING equilateral triangle symbol appears throughout 
the manual accompanying certain maintenance and repair activities.  
Please follow the important instructions accompanying this symbol 
to avoid situations that could cause injury to the operator or other 
personnel.   

CAUTION 

WARNING 

Summary of Contents for 830-00905

Page 1: ...Recirculating Chiller Model 830 00905 830 00906 Operation Manual Sonics Materials Inc 53 Church Hill Road Newtown CT 06470 USA Phone 203 270 4600 info sonics com www sonics com Rev 12 21...

Page 2: ..._________________________ 8 START UP__________________________________________________________________________ 8 SECTION 5 _________________________________________________________________________ 8 O...

Page 3: ...ets the provisions of the directives 2004 108 EC EMC Directive 2006 95 EC Low Voltage Directive EN 61326 1 2006 Emissions and Immunity EN 61000 3 2 2006 Harmonics Emissions EN 61000 3 3 2008 Voltage F...

Page 4: ...ng point Only water is recommended when using a 830 00905 with an ultrasonic processor and cup horn microplate horn accessory When using a 830 00906 always use only proper coolants as specified in man...

Page 5: ...s The system provides 1 liter per minute of constant temperature coolant with PID control for both cooling and heating With fewer moving parts the system is highly reliable and energy efficient The ch...

Page 6: ...00906 DI Water Koolance 27 propylene glycol water mix or 27 50 ethylene glycol water mix Process Fluid Fittings 1 4 quick connect type connectors Colder brand Pumps Diaphragm Wetted Materials Aluminu...

Page 7: ...830 00905 830 00906 Side Figure 3C 830 00905 830 00906 Back Filter Wrench used to tighten or remove filter housing Process Fluid Connectors On Off Switch Signal Connector 9 Pin D sub Display Screen Co...

Page 8: ...20mm GDB quick acting glass meets IEC 127 2 Replacement Fuse Sonics 20 22332 10 Allied Electronics 70149445 Optional Alarms Alarm signals are TTL signals normally high 4 VDC located on the 9 pin d sub...

Page 9: ...nue using the filter on the back of the unit Connect the Coolant Return tubing from the flow cell water jacket directly to the chiller Coolant Return port bypassing the filter Recommended Coolants DI...

Page 10: ...LEVEL LOW add more coolant to the reservoir until the display changes to read the coolant temperature Important Notes Do not run the chiller dry for more than a few seconds SECTION 5 OPERATION _______...

Page 11: ...to the set point temperature To change the set point temperatures just press the UP or DOWN keys again to change SETTEMP 1 to the new set point followed by ENTER and then START STOP The Operating Mode...

Page 12: ...lly TTL low 0 5 VDC system failure has occurred A list of system failures causing the system alarm to change to TTL low can be found in Section 6 In the event of a system failure the alarm type will b...

Page 13: ...for a RMA number to return the unit for fan replacement Pump Fail The liquid heat exchanger plate temperature is either too hot or too cold indicating pump failure a blockage in the external plumbing...

Page 14: ...larms are present poor temperature control indicates the heat load is too great for the chiller the TE cooling heating engine is not receiving power the PID constants have been corrupted or the chille...

Page 15: ...non shedding wipe dipped in isopropyl alcohol SECTION 9 TECHNICAL SUPPORT _____________________________________________ Customer service is our highest priority Please contact us immediately for tech...

Page 16: ...ich invalidate the warranty include but are not limited to Removing the serial number label Any disassembly partial or complete of the product Changing any components of the product Subjecting the pro...