Sonics 830-00905 Operation Manual Download Page 10

 

 

 

R

EV

.

  

12-21

 

S

ECTION 

S

TART 

U

  ______________________________________________________ 

 

  

 

Note: In order to avoid injury or damage, operators must only use this product in the 

manner specified below. 

 
Start-up the chiller using the following steps: 

 

1)

 

Connect coolant tubing to fluid connections located on the left 
side  of  the  unit,  labeled  Process  Out  (supply)  and  Process  In 
(return).  

2)

 

When  using  a  #830-00905  chiller:  review  the  Cup  horn  or 
Microplate  horn  accessory  instructions for information on how 
to fill with DI water. 
When using a #830-00906 chiller: Remove the reservoir cap on 
top  and  fill  the  reservoir  to  just  below  the  bottom  of  its  neck 
with coolant.  Replace cap. 

3)

 

Plug line cord into 115 - 230 VAC, 50/60 Hz. 

4)

 

Turn  on  switch  located  on  the  left  side  of  the  unit.    The front 
display should read the current coolant temperature.   
Note:  When  using  the  #830-00906  chiller:  If  the  front  display 
reads “TANK LEVEL LOW”, add more coolant to the reservoir 
until the display changes to read the coolant temperature. 

 
Important Notes:  
 

Do not run the chiller dry for more than a few seconds. 

 

   
 
 

 

S

ECTION 

O

PERATION

  _____________________________________________________  

 
The  chiller  is  operated  via  the  control  panel  located  on  the  front 
panel.  The control panel has a 16-character LCD display and four 
input keys: UP, DOWN, ENTER, and START/STOP.  These keys 
work as follows: 

 

Key 

Action 

UP 

Pressing the UP key raises the parameter value displayed.  

DOWN 

Pressing the DOWN key lowers the parameter value displayed 

ENTER 

Pressing the ENTER key momentarily enters the parameter changed. 

ENTER 

Pressing  and  holding  the  ENTER  key  for  3  seconds  causes  the  chiller  to  change  the 
display menu (see menu structure) 

START/STOP  Pressing the START/STOP key turns on temperature control. 
START/STOP  Pressing  the  START/STOP  key  while  the  chiller  is  operating  turns  off  temperature 

control (Operating Mode = *). 

WARNING 

Electrical Shock 

Hazard:  Never Plug 

in a Line Cord with 

Wet Hands 

Running the     

ThermoCube dry  

(no fluid) may  

damage the pump 

CAUTION 

Summary of Contents for 830-00905

Page 1: ...Recirculating Chiller Model 830 00905 830 00906 Operation Manual Sonics Materials Inc 53 Church Hill Road Newtown CT 06470 USA Phone 203 270 4600 info sonics com www sonics com Rev 12 21...

Page 2: ..._________________________ 8 START UP__________________________________________________________________________ 8 SECTION 5 _________________________________________________________________________ 8 O...

Page 3: ...ets the provisions of the directives 2004 108 EC EMC Directive 2006 95 EC Low Voltage Directive EN 61326 1 2006 Emissions and Immunity EN 61000 3 2 2006 Harmonics Emissions EN 61000 3 3 2008 Voltage F...

Page 4: ...ng point Only water is recommended when using a 830 00905 with an ultrasonic processor and cup horn microplate horn accessory When using a 830 00906 always use only proper coolants as specified in man...

Page 5: ...s The system provides 1 liter per minute of constant temperature coolant with PID control for both cooling and heating With fewer moving parts the system is highly reliable and energy efficient The ch...

Page 6: ...00906 DI Water Koolance 27 propylene glycol water mix or 27 50 ethylene glycol water mix Process Fluid Fittings 1 4 quick connect type connectors Colder brand Pumps Diaphragm Wetted Materials Aluminu...

Page 7: ...830 00905 830 00906 Side Figure 3C 830 00905 830 00906 Back Filter Wrench used to tighten or remove filter housing Process Fluid Connectors On Off Switch Signal Connector 9 Pin D sub Display Screen Co...

Page 8: ...20mm GDB quick acting glass meets IEC 127 2 Replacement Fuse Sonics 20 22332 10 Allied Electronics 70149445 Optional Alarms Alarm signals are TTL signals normally high 4 VDC located on the 9 pin d sub...

Page 9: ...nue using the filter on the back of the unit Connect the Coolant Return tubing from the flow cell water jacket directly to the chiller Coolant Return port bypassing the filter Recommended Coolants DI...

Page 10: ...LEVEL LOW add more coolant to the reservoir until the display changes to read the coolant temperature Important Notes Do not run the chiller dry for more than a few seconds SECTION 5 OPERATION _______...

Page 11: ...to the set point temperature To change the set point temperatures just press the UP or DOWN keys again to change SETTEMP 1 to the new set point followed by ENTER and then START STOP The Operating Mode...

Page 12: ...lly TTL low 0 5 VDC system failure has occurred A list of system failures causing the system alarm to change to TTL low can be found in Section 6 In the event of a system failure the alarm type will b...

Page 13: ...for a RMA number to return the unit for fan replacement Pump Fail The liquid heat exchanger plate temperature is either too hot or too cold indicating pump failure a blockage in the external plumbing...

Page 14: ...larms are present poor temperature control indicates the heat load is too great for the chiller the TE cooling heating engine is not receiving power the PID constants have been corrupted or the chille...

Page 15: ...non shedding wipe dipped in isopropyl alcohol SECTION 9 TECHNICAL SUPPORT _____________________________________________ Customer service is our highest priority Please contact us immediately for tech...

Page 16: ...ich invalidate the warranty include but are not limited to Removing the serial number label Any disassembly partial or complete of the product Changing any components of the product Subjecting the pro...

Reviews: