OPERATING MANUAL PAT-810/815/820 version 1.06
80
adjustable measuring time in the range of: 1 s
…60
s with resolution of 1 s and option
Con-
tinuous test
Measurement of PE leakage current and differentia current with clamp
Display range
Resolution
Measurement uncertainty
(basic)
0,00 mA…9,99 mA
0,01 mA
(5 % m.v. + 5 digits)
10,0 mA…99,9 mA
0,1 mA
Basic uncertainty does not include the uncertainty of the current clamp
adjustable upper limit in the range of
Ś 0,01 mA…19,90 mA
with resolution of 0,01 mA
adjustable measuri
ng time in the range ofŚ 1…180s with resolution of 1 s and option
Contin-
uous test
Measurement of touch leakage current
Note:
The tester changes polarity automatically in the mains test socket during test, and it displays
higher measured value as the final result.
Display range
Resolution
Measurement uncertainty
(basic)
0,000 mA...4,999 mA
0,001 mA
(5 % m.v. + 3 digits)
Influencing factor
Designation
Additional uncertainty
Position
E
1
0 %
Supply voltage
E
2
0 %
Temperature
E
3
0,25
μA/ºC
The shape of the network
voltage (CF)
E
8
0 %
the bandwidth of test current results from the measuring system with adjusted touch current
which simulates human perception and reaction, in accordance with EN 60990: 2002
adjustable upper limit in the range of: 0.01 mA... 1.99 mA with resolution 0,01 mA
adjustable measuring time in the range of: 1 s
…60
s with resolution of 1 s and option
Continu-
ous test
S Power Measurement
Display range
Resolution
Measurement uncertainty
(basic)*
0 VA...999 VA
1 VA
(5 % m.v. + 3 digits)
1 kVA…3,99 kVA
0,01 kVA
adjustable measuring time
in the range of: 1 s...60 s, with resolution of 1 s and option
Continuous test
(enabled by default), in AUTOTEST only adjustable: 1...60 s, with resolution
of 1 s
* only for current measurement with clamp
(8 % m.v. + 5 digits)
P Power measurement
Display range
Resolution
Basic uncertainty*
0 VA...999 W
1 W
(5 % m.v. + 3 digits)
1 kVA…3,99 kW
0,01 kW