OPERATING MANUAL PAT-810/815/820 version 1.06
54
Negative test result: I
L
> LIMIT
Note:
During the measurement in the test socket the mains voltage is present.
4.11 IEC cable test
IEC lead test includes wires continuity checking, short circuits between the wires and correctness
of L-L and N-N connection and PE resistance and insulation resistance measurement.
Press
Test
IEC
.
Before the measurement, in
the same manner as in previ-
ous measurements, the fol-
lowing parameters must be
set: current, measurement
time and limit R
PE
as well as
measurement time and limit
R
ISO
.
Connect the mains plug of the cord into the test socket and the other end of the cord into the
IEC socket on the tester.