OPERATING MANUAL PAT-810/815/820 version 1.06
78
detection of a dangerous voltage before commencing a measurement
discharging the object tested
Note:
For R < 50 k
Ω
, the uncertainty is not specified.
Measurement of insulation resistance using test voltage of 500V
Test range according to IEC 61557-2 for U
N
= 500 V: 500 k
Ω…599.9
M
Ω
Display range
Resolution
Measurement uncertainty
(basic)
0 k
Ω
...1999 k
Ω
1 k
Ω
(5 % m.v. + 8 digits)
2,00 M
Ω
...19,99 M
Ω
0,01 M
Ω
20,0 M
Ω
...599,9 M
Ω
0,1 M
Ω
Influencing factor
Designation
Additional uncertainty
Position
E
1
0 %
Supply voltage
E
2
0 %
Temperature
E
3
0,1 %/ºC
Capacity
E
7
0 % for R
≤
20 M
Ω
Unspeci
fied for R > 20 MΩ
Accuracy of generated voltage (Robc [
]
1000*U
N
[V]): -0 % +30 % from the set value
nominal current: min 1 mA...1.4 mA
adjustable lower limit within the range of 0.1 M
Ω
….9.9 MΩ
with resolution of 0,1 M
Ω
adjustable measuring time: 3
s…3
min with a resolution of 1 s and option
Continuous test
detection of a dangerous voltage before commencing a measurement
discharging the object tested
Note:
For R <50 k
Ω
, the uncertainty is not specified.
Flash test (only PAT-820)
Display range
Resolution
Measurement uncertainty
(basic)
0,00 mA...2,50 mA
0,01 mA
(5 % w.m. + 5 cyfr)
measurement voltage: 1500 V AC, 3000 V AC
adjustable measuring time
Ś 2 s…180 s
adjustable upper limit in the range of
Ś 0,01 mA…2,50 mA
with resolution
0,01 mA / 0,1 mA
Measurement of substitute leakage current
Display range
Resolution
Measurement uncertainty
(basic)
0,00 mA...3,99 mA
0,01 mA
(5 % m.v. + 2 digits)
4,0 mA...19,9 mA
0,1 mA
Influencing factor
Designation
Additional uncertainty
Position
E
1
0 %
Supply voltage
E
2
0 %
Temperature
E
3
0,075 %/ºC
opening voltage: 25 V...50 V
internal resistance of the testing device 2 k
Ω
± 20
%
adjustable upper limit in the range of: 0.01 mA... 9.90 mA with resolution of 0.01 mA