MRU-30
– USER MANUAL
28
4.9 Calibration of the test leads for R
CONT
measurement
To eliminate the influence of the resistance of the test leads on the result of
R
CONT
measurement
perform this compensation (auto-zeroing) procedure while in
R
CONT
measurement mode.
In
R
CONT
mode LED
is illuminated press
SET/SEL
to
switch into auto-zeroing mode.
The flashing
AUTO-ZERO
mes-
sage indicates the test leads cali-
bration procedure is ready to
begin.
Press the
START
button.
When a steady
AUTO-ZERO
message is displayed on the
screen, the test leads calibration
procedure is complete.
The results now include the cor-
rection value for
R
CONT
and is indi-
cated by
AUTO-ZERO
displayed
on screen. This value is saved in
memory and will be overwritten
whenever the Auto-Zero proce-
dure is performed
Connect the test leads according to this diagram:
Short-circuit the test leads connected to the
‘E’ and ‘S’ terminals.