Somat eDAQ
lite
82
I2773-4.
4
en
Show Grids
Select the show grids option to show the grid lines on the display.
4.6
Uploading Test Data
Uploading test data transfers all or selected test runs resident in the data file stored in
the eDAQ
lite
to a user-specified PC disk file. Upload complete SIE or SIF files after
stopping a test run. When using SIE data, partial files are available for upload during
a test run.
After transferring the file to the support PC, use InField or other data analysis software
to view and analyze the test data.
4.6.1
Uploading SIE Data Files
From the eDAQ
lite
To upload SIE data from the RAM disk or PC Card storage in the eDAQ
lite
, select
Upload Test Data from the Test Control menu or toolbar. Upload SIE test data at any
time, including during a test run when partial test data is available. Use some caution
when uploading during a test run as the upload task consumes some processing time
and can result in a DeviceOverFlow error reset if the eDAQ
lite
cannot keep up with
the processing requirements.
4.6.2
Uploading SIF Data Files
From the eDAQ
lite
To upload SIF data from the RAM disk or PC Card storage in the eDAQ
lite
, select
Upload Test Data from the Test Control menu or toolbar. If using RAM disk storage,
TCE uploads the SIF file directly to the PC file. If the PC Card data storage option is
in use, the eDAQ
lite
first consolidates the SIF data file from the RAM disk resident
component file and the set of PC Card resident component files. A test run cannot be
in progress to upload a SIF data file.
Upload SIF data after a test run or after ending a test. The upload option becomes
unavailable as soon as a new test is initialized, since the eDAQ
lite
deletes the RAM
disk SIF component file upon initialization.
NOTE
If there is a communications failure between the PC and the eDAQ
lite
during an
upload of the files on the PC Card, TCE enters a loop of making repeated attempts to
re-establish communications so that the upload process can recover and continue.
This recovery scheme handles most types of communication failures such as power
Plot Mode
Description
Normal
When the plot reaches the end of the x-axis, TCE erases the display
and draws the next vertical bar at the beginning of the x-axis.
Scroll
The vertical bars continuously scroll from left to right after the plot
reaches the end of the x-axis for the first time. The latest min and max
values are always at the end of the x-axis.
Mixed
When the plot reaches the end of the x-axis, TCE moves the last half
of the display to the first half and resumes plotting from the middle of
the display.
HBM: public
Summary of Contents for eDAQlite
Page 1: ...User Manual English Deutsch Espanol I2773 4 4 en HBM public Somat eDAQlite with TCE Software ...
Page 2: ...Somat eDAQlite 2 I2773 4 4 en HBM public ...
Page 16: ...Somat eDAQlite 16 I2773 4 4 en HBM public ...
Page 22: ...Somat eDAQlite 22 I2773 4 4 en HBM public ...
Page 88: ...Somat eDAQlite 88 I2773 4 4 en HBM public ...
Page 106: ...Somat eDAQlite 106 I2773 4 4 en HBM public ...
Page 128: ...Somat eDAQlite 128 I2773 4 4 en HBM public ...
Page 190: ...Somat eDAQlite 190 I2773 4 4 en HBM public ...
Page 202: ...Somat eDAQlite 202 I2773 4 4 en HBM public ...
Page 222: ...Somat eDAQlite 222 I2773 4 4 en HBM public ...
Page 224: ...Somat eDAQlite 224 I2773 4 4 en HBM public ...