Drift Detect
Select the drift detect routine to enable drift detection. TCE sets the reference mean
as the mean value of the first window for each test run. If the difference between the
current window mean and the reference mean exceeds the specified mean gate, bit 2
of the output byte is set to 1.
Limit Detect
Select the limit detect routine to enable limit detection. If any data sample in the
analysis window is greater than the specified maximum limit or less than the specified
minimum limit, bit 3 of the output byte is set to 1.
Kurtosis Detect
Select the kurtosis detect routine to enable kurtosis detection. If the kurtosis
coefficient for the data in the analysis window is greater than the specified maximum
limit, then bit 4 of the output byte is set to 1. The eDAQ
lite
uses the following equation
to calculate the kurtosis coefficient:
where,
N
is the number of data samples in the analysis window and
x
mean
is the
mean of the data samples in the analysis window.
7.5.5
Triggered Latch
The Triggered Latch channel generates an output channel that latches the previous
input channel data value when triggered. By default, the channel latches the input data
value when the trigger channel transitions from FALSE to TRUE and holds this data
value until the trigger channel transitions from TRUE to FALSE. When the trigger
channel is FALSE, the channel outputs the same value as the input channel.
NOTE
If the trigger channel is initially TRUE, the channel outputs a system defined value until
the trigger channel transitions to FALSE. The defined value is based on the input
channel data type.
Input Channel
The input channel data type must be 32-bit float, 32-bit integer or 32-bit unsigned.
Output Data Type
The output data type is the same as the input data type.
Trigger Channel
Select the input channel used to determine when to latch the input channel.
Input Data Type
System Defined Value
32-bit float
0.0
32-bit integer
Actual full-scale mean value (typically 0.0)
32-bit unsigned
Actual full-scale minimum value (typically 0.0)
HBM: public
Summary of Contents for eDAQlite
Page 1: ...User Manual English Deutsch Espanol I2773 4 4 en HBM public Somat eDAQlite with TCE Software ...
Page 2: ...Somat eDAQlite 2 I2773 4 4 en HBM public ...
Page 16: ...Somat eDAQlite 16 I2773 4 4 en HBM public ...
Page 22: ...Somat eDAQlite 22 I2773 4 4 en HBM public ...
Page 88: ...Somat eDAQlite 88 I2773 4 4 en HBM public ...
Page 106: ...Somat eDAQlite 106 I2773 4 4 en HBM public ...
Page 128: ...Somat eDAQlite 128 I2773 4 4 en HBM public ...
Page 190: ...Somat eDAQlite 190 I2773 4 4 en HBM public ...
Page 202: ...Somat eDAQlite 202 I2773 4 4 en HBM public ...
Page 222: ...Somat eDAQlite 222 I2773 4 4 en HBM public ...
Page 224: ...Somat eDAQlite 224 I2773 4 4 en HBM public ...