13 TIMING AND MEASUREMENT INFORMATION
Parameter
Update rate
Comment
Measurement Reading Rate
5 mS
10mS for SI400
Digital Output Update
5 mS from measurement
10mS for SI400
Response to Digital Input
5 mS
10mS for SI400
Probe Reading Rate
256uS
The probe connected to the ACS unit and
any external digital probes are updated at
this rate
Averaging is applied at this rate. i.e. 16 x
averaging will complete every 4.096mS
Probe Averaging
1,2,4,8,16,32,64,128,256 cycles
Averaging is applied at this rate. i.e. 16 x
averaging will complete every 4.096mS
Measurement Bandwidth (electronics)
4,8,16,32,64,128,256, 512 Hz
Measurement bandwidth directly related to
the Probe averaging value
1 – 512Hz
256 - 4Hz
Resolution
14 bits
16 bits default
18 bits
To calculate the resolution for a probe
1 bit = Probe Range (microns)/2^
resolution
503110 - Orbit ACS Manual
Issue 9
38 of 63
Summary of Contents for AMETEK ORBIT ACS Series
Page 1: ......
Page 17: ...9 1 SI100 AND SI200 DISPLAYS 503110 Orbit ACS Manual Issue 9 17 of 63 ...
Page 23: ...9 3 1 Top Level Menu 503110 Orbit ACS Manual Issue 9 23 of 63 ...
Page 24: ...9 3 2 Configure Probe s 503110 Orbit ACS Manual Issue 9 24 of 63 ...
Page 29: ...9 3 5 Peak Preset Limits and Misc Menu 503110 Orbit ACS Manual Issue 9 29 of 63 ...
Page 30: ...9 3 6 Output Input and Communications Menu 503110 Orbit ACS Manual Issue 9 30 of 63 ...
Page 47: ...15 5 ASCII CHARACTER SET Example ASCII A is HEX 41 503110 Orbit ACS Manual Issue 9 47 of 63 ...