Sofraser MIVI 9610 Technical Manual Download Page 5

 

Technical Manual MIVI 9612

 

 

Réf. 254-31/0 

 

1.

 

General presentation 

The measuring chain is composed of two indissociable elements : the sensor and the electronic 

processor. 

1.1. The sensor 

Each  sensor  is  matched  with  its  own  electronics.  Before  switching  on,  be  sure  that  serial 

numbers at the sensor and the electronics identification plates are corresponding. 

The  operation  principle  of  the  equipment  makes  that  the  provided  viscosity  information  is 

relative : in the same fluid, at the same conditions the information is the same. But for 2 rheologically 
different fluids the response can be different (despite it is perfectly repeatable). 

The sensor active part is protected by an stainless tube which must not be removed. 

The sensor is fixed upon its mounting flange by means of the provided 4 long screws. 

See the detailed leaflet, § 2. 

1.2. The processor 

Its functions: 

Powers the sensor, 

Processes the sensor’s information, 

Displays  and  transmits  the  process  information  :  viscosity,  temperature, 

sensor’s frequency, outputs and relays states…. 

The processor is composed of two modes : 

“UTILISATION” Mode : 
 

The  “UTILISATION”  Mode  makes  it  possible  to  visualize  in  different  forms  viscosity  and 

temperature values. 
 

Visualization with bargraphs, acquisition graph 

Access to the sensor identification data 

Access to the “ADAPTATION” Mode blocks starting from the principal menu 

 
“ADAPTATION” Mode : 
 

The “ADAPTATION” Mode makes it possible to adapt the displayed data, parameters, settings 

and configuration of the processor according to the using conditions. 
 

Display  block :  makes  it  possible  to  display  outputs  and  relays  states,  rough  data,  to 
choose viscosity and temperature units and to choose the viscosity correction to active. 

Settings  block :  makes  it  possible  to  set  current  outputs,  alarms,  bargraphs  and  graph 
scales, to choose and set the modalities parameters of kinematic viscosity calculation, to 
proceed  to  the  mounting  offset  adjustment,  to  activate  and  define  viscosity  filter  and  to 
change the Pt100 status. 

Parameters  block :  makes  it  possible  to  parameter  viscosity  corrections,  to  read  and 
modify the compensation table, to modify security codes and to set the density coefficient. 

Configuration block : makes it possible to SOFRASER to configure the processor with the 
manufacturing data (thermal drift, calibration) and to define the viscosity adjustment data. 

 

Summary of Contents for MIVI 9610

Page 1: ...Quality System certified instruments 9612 Version Technical Manual MIVI 9610 R f 254 31 0...

Page 2: ...THE ROD MUST VIBRATE IN AIR 3 INSTALL THE SENSOR ON THE PROCESS AND FIX IT WITH ITS 4 SCREWS 4 REACH THE OFFSET ADJUSTMENT FUNCTION STARTING FROM THE BLOCK SETTINGS AND WITH THE SECURITY CODE 2 5 FOL...

Page 3: ...2 2 Plane side mounting 10 2 2 3 Pot mounting for small flow rates or pilot plant 10 2 2 4 Special mountings 10 2 2 5 Replacement cap 10 2 3 PRACTICAL ADVICES 11 2 4 CHECKING 11 2 5 SENSOR WIRING 11 2...

Page 4: ...E 23 6 2 RS485 INTERFACE 23 7 POWER SUPPLY 25 7 1 CHARACTERISTICS 25 7 2 SAFETY CONSIDERATIONS 25 7 3 POWER SUPPLY WIRING 26 8 GENERAL CABLING SCHEMATICS 27 9 SPECIFIC NOTES AND MANUFACTURING PARAMETE...

Page 5: ...es UTILISATION Mode The UTILISATION Mode makes it possible to visualize in different forms viscosity and temperature values Visualization with bargraphs acquisition graph Access to the sensor identifi...

Page 6: ...re The rod is vibrating in air When possible note the viscosity information when a cleaning or rinsing solution is flowing If the original calibration is convenient one of the 2 above mentioned values...

Page 7: ...gas or II 2D dust Ex d IIC T1 to T6 gas Ex tD A21 IP67 T75 C T90 C T125 C T190 C T290 C T300 C dust Be sure the sensor s certification is in accordance with the security level required on your proces...

Page 8: ...lded on the process line It means it can not be individually certified but with the whole process line 1 6 4 EMC and low voltage directive The processor 9601 is in agreement with EMC specifications de...

Page 9: ...of the network or reactor If the fluid temperature varies widely and fast choose the upside down or horizontal position in order to allow a convenient air convection among the sensor body It is screw...

Page 10: ...order to avoid parasitic vibrations the plate where the flange is welded must be thicker than 5 mm Advice preferably choose a sensor position with the rod to the top in order to have a total immersio...

Page 11: ...to bend the vibrating rod A ring with the same dimensions of the protector s base must be placed on the head of the sensor in order to maintain the O ring IMPORTANT As soon as the sensor is removed s...

Page 12: ...2 Standard and explosion proof models are both equipped with cable glands Those ones can match the ATEX norm depending on the MIVI version Earth terminal Pic 4 Shielding Cable Receiving coil Driving c...

Page 13: ...NERAL PURPOSE Ex proof 200 C Std 200 C 135 mm 120 mm 100 mm 17 mm 30 mm Guard tube short slits Pot gas 156 mm 156 mm HIGH PRESSURE HP 200 C HP ex proof 200 C Pot gas Guard tube long slits Low pressure...

Page 14: ...Technical Manual MIVI 9612 R f 254 30 0 14 2 7 Tightness O ring 29 82 x 2 62 76 or 79 70 or 99 O ring 21 95 x 1 78 O ring 30 00 x 2 00 O ring 30 00 x 2 50 Pic 6...

Page 15: ...l dimensions Height 80mm Width 100mm Length 160mm Tightness IP66 Material Aluminium Temperature from 0 C to 45 C 3 2 Transducer s electronic board wiring The transducer s board is wired through the co...

Page 16: ...26 mm Thickness 5mm Weight about 515 g Tightness IP 65 with panel mounting and IP20 on rear panel Operational temperature 0 C to 45 C 4 2 Display Display type STN LCD Illumination backlight LED yellow...

Page 17: ...ply wiring 4 7 Connection blocks On the processor back module various categories are identified example Digital Inputs Analog Outputs We will call these categories connection blocks The diagram below...

Page 18: ...oltage 0 10 V current 0 20 mA or 4 20 mA These inputs can be used with thermocouple or Pt100 On 9612 electronics these inputs are connected as 4 20mA and are specific to the plug of a densimeter Four...

Page 19: ...om 02 to 07 and from 08 to 013 3 A max 250VAC or 30VDC Common 8 A max common plug not written see the connecting plan Minimum load 1mA 5VDC Lifespan 100 000 operations with a maximum load Response tim...

Page 20: ...in If the user does not have the means to take in charge the wastes of our electronic equipments he should returned these equipments to our factory with a signed letter confirming that it is an equipm...

Page 21: ...hole for the provided NC6 32 screw Scrape the panel s paint away to ensure the conductive connection Drive the screw into the hole and place the following items first washer ring cable shoe second was...

Page 22: ...254 30 0 22 Step 4 Push the 4 mounting brackets into their slots on the processor sides shown as follow Step 5 Tighten the bracket screws on the panel Firmly hold the mounting brackets on the process...

Page 23: ...on 4 RxD signal reception 5 0V reference 6 DSR signal 6 2 RS485 interface RS485 port is used to make the connection between the transmitter and the device Used COM port COM 3 Serial port type RJ 11 Ca...

Page 24: ...munication capacity up to 32 devices Voltage limits from 7 V to 12 V as max differential Pin number Function 1 A signal red terminal 2 3 4 5 6 B signal white terminal Pins from 2 to 5 are not used for...

Page 25: ...upply can be used provided that a 0V signal is connected to the frame If the device is mounted on a metal panel standard safety considerations require that the power supply must be earthed to avoid el...

Page 26: ...the wire do not exceed a maximum torque of 0 5 N m We recommend that you use crimp terminals for wiring Use 26 14 AWG wire for all wiring purposes To maximize the system s performance electromagnetic...

Page 27: ...cted to the 9612 device Pieces of advice The transducers must all have the same power supply Put a 120 resistance at the end of the line 1 4 W power is enough The RS485 shielding has to be connected t...

Page 28: ...Technical Manual MIVI 9612 R f 254 30 0 28...

Page 29: ...erature LOW 0 0 HIGH 999 9 HYST 0 0 HYST 0 0 Alarms attribution CURRENT OUTPUTS Physical value Current output value Viscosity LOW 0 0 4 HIGH 999999 9 20 BURN 4 Temperature LOW 0 0 4 HIGH 999 9 20 BURN...

Page 30: ...999999 9 999999 9 Line 7 999999 9 999999 9 Line 8 999999 9 999999 9 Line 9 999999 9 999999 9 COMPENSATION TABLE X Y Line 0 0 0 0 0 Line 1 999 9 0 0 Line 2 999 9 0 0 Line 3 999 9 0 0 Line 4 999 9 0 0 L...

Page 31: ...1 999 0 Line 2 999 0 Line 3 999 0 Line 4 999 0 Line 5 999 0 Line 6 999 0 Line 7 999 0 Line 8 999 0 Line 9 999 0 VISCOSITY ADJUSTMENT A 1 000 B 0 000 CALIBRATION DATA SERIAL N Y0 0 0000 a 1 0000 b 1 00...

Reviews: