Technical Manual MIVI 9612
Réf. 254-31/0
2
IMPORTANT
BEFORE STARTING THE PROCESS, IT IS VERY
IMPORTANT TO PROCEED TO THE OFFSET
ADJUSTMENT IN AIR:
The procedure of the offset adjustment is detailed in the user manual
§ 5.1:
1.
CLEAN AND DRY THE SENSOR ROD.
2.
BE SURE THE PROCESS IS EMPTY. THE ROD MUST VIBRATE IN AIR.
3.
INSTALL THE SENSOR ON THE PROCESS AND FIX IT WITH ITS 4 SCREWS.
4. REACH
THE
OFFSET
ADJUSTMENT
FUNCTION
STARTING
FROM
THE
BLOCK “SETTINGS” AND WITH THE SECURITY CODE 2.
5. FOLLOW THE INSTRUCTIONS DISPLAYED ON THE ELECTRONICS SCREEN.
6. PRESS ON “F3” AND “OK” TO ADJUST THE OFFSET. IT MEANS THAT V
BRUT
WILL BE
SHIFTED TO THE REFERENCE 900 mV.
PRESSING ON “OK” IS POSSIBLE ONLY IF V
BRUT
IS WELL SHIFTED TO 900 mV.
A MESSAGE SAYING THE OPERATION WAS SUCCESSFUL APPEARS.