Settings and Calibration
1. Press and hold any key while turning the scale ON.
CAL 1
is shown on the LCD display.
key to toggle among the
CAL 1
,
CAL 2
,
CAL 3
,
and CAL 4
menus.
2. Press
3. Select
CAL 1
, press
to enter General Settings menu.
4. Select
CAL 2
, press
to enter the Print Settings menu
6. Select
CAL 3
, press
to enter Calibration procedure.
7. Select
CAL 4
, press
to enter Linearity Calibration procedure.
Note: To perform CAL 3 and CAL 4 calibration, you must remove the JP3 jumper from the PCB first.
Put the JP3 back after calibration is complete.
General Settings (CAL 1)
to toggle among the options, and
In the General Settings menu, press
to confirm.
1. Aut.oFF– Auto Off Setting
Display Descriptions
A.oFF 0
Disable auto shutdown function.
A.oFF 1
If there is no operation, the scale will shut down in 5 minutes.
A.oFF 2
If there is no operation, the scale will shut down in 10 minutes.
A.oFF 3
If there is no operation, the scale will shut down in 20 minutes.
A.oFF 4
If there is no operation, the scale will shut down in 30 minutes.
2. Pcnt.dP – Percentage decimal Setting
Display Descriptions
Pcnt.dP 0.0
100.0 %.
Pcnt.dP 0.00
100.00 %
3. bbH - Hi Beep Setting
Display Explanation
bbH_ 0
Disable Hi alarm sound.
bbH_ 1
Set Hi alarm sound as continual short beeps.
bbH_ 2
Set Hi alarm sound as continual long beeps.
8