Snowrex GW-15D User Manual Download Page 4

Display and keypad 

LCD Display 

 

Numerical display to display weight, ID number 

 

 

3

 

Indicates the weight is stable.

 

Indicates the weight is at zero.

 

 

MR 

Indicates the total weights in memory 

M+

 

Indicates the accumulation weights in memory.

 

G.W

 

When 

 is shown above the G.W. indicator, the Gross 

Weight is displayed.

 

NET 

When 

 is shown above the NET. indicator, the Net 

Weight is displayed.

 

T.W 

When 

 is shown above the T.W. indicator, the Tare 

Weight is displayed.

 

Hi 

When the weight is over the weight upper bound, 

 is 

flashing next to the Hi indicator.

 

Go 

When the weight is between the weight upper bound and 
lower bound, 

 is flashing next to the Go indicator. 

Lo 

When the weight is under the weight lower bound, 

 is 

flashing next to the Lo indicator.

 

 is shown above the 

  when in counting mode. 

 

 is shown above the 

  when in percentage mode. 

 

kg & lb 

Indicate the unit of weight 

 

Summary of Contents for GW-15D

Page 1: ...GW Professional Weighing Scale User Manual SNOWREX INTERNATIONAL CO LTD SRGW 20100415...

Page 2: ...and keypad 3 LCD Display 3 Keypad 4 Using the weighing scale 6 Counting function 6 Percentage function 6 Operation of M 6 Operation of MR 6 Alarm function 7 Settings and Calibration 8 General Settings...

Page 3: ...0mA AC adapter Built in 6V Rechargeable Battery Interface RS 232C optional Series specification EC TYPE OIML APPROVED Model GW 6D GW 15D GW 30D Max1 Max2 3kg 6kg 6kg 15kg 15kg 30kg Min1 Min2 20g 3kg 4...

Page 4: ...bove the NET indicator the Net Weight is displayed T W When is shown above the T W indicator the Tare Weight is displayed Hi When the weight is over the weight upper bound is flashing next to the Hi i...

Page 5: ...ut data M key Press to accumulate piece counting and weighing results MR key Press to recall total number of accumulation and total weighing Alarm key Press to set alarm for weight upper and lower bou...

Page 6: ...to switch between kg and lb ID key Press to enter the ID mode Use the numeric keypad to enter the user s ID number Gross Net Tare key Press to toggle among Gross Weight Net Weight and Tare Weight on...

Page 7: ...ghing pan press and after a beep sound the indicator let to M icon will light up on the LCD indicating a data has been recorded 2 Clear the load and put another load on the weighing pan When the scale...

Page 8: ...Hi setting And then move to Lo setting 3 The Lo is showing in the left window Press the keys to set the lower bound of the weight 4 Press to confirm to save your Lo setting And then back to weighing...

Page 9: ...ral Settings CAL 1 to toggle among the options and In the General Settings menu press to confirm 1 Aut oFF Auto Off Setting Display Descriptions A oFF 0 Disable auto shutdown function A oFF 1 If there...

Page 10: ...criptions Pr dEFA Set the print format to default settings Pr LP50 Set the print format to LP 50 printer 2 Printout setting Display Descriptions P PrESS Press M to print single weighing result Press C...

Page 11: ...ed for sending weight data to PC P oFF Print function disabled 3 Baud Rate Display Descriptions br2400 Set the baud rate to 2400 br4800 Set the baud rate to 4800 br9600 Set the baud rate to 9600 4 Par...

Page 12: ...t Calibration Display Descriptions The right window shows AD Value of Zero weight Wait until the AD value becomes stable press Pnt 0 XXXXXX to confirm 4 Setup calibration weight Display Descriptions P...

Page 13: ...m capacity Display Descriptions The AD Value of 1 3 of the full load is then displayed Pnt 1 Put on poise of the 1 3 weight by the maximum capacity Wait until the AD value becomes stable press to proc...

Page 14: ...cates it s an unstable weight S 53h indicates it s a stable weight SIGN 2Dh means it s a negative weight 20h space char indicates it s a positive or zero weight W6 W0 Weight data decimal point include...

Reviews: