Appendix #2: Resin Cleaning Procedure/TOC in the Feed Stream
The organic cleaner may either be an approved
non-ionic
surfactant solution, such
as Triton X-100
tm
, or an approved caustic-based solution, such as Avista RoClean
tm
L211.
Recommendation
: Minimize the organic fouling of the resins in the EDI system by
minimizing TOC (Total Organic Carbon) via the RO and its pretreatment system.
Suggestion
: Use 3-way valves in the design of a CIP system to eliminate the need to
manually disconnect the plumbing when servicing the EDI module.
Surfactant Solution:
10 liters deionized water
10 ml of pure Triton X-100
tm
(or 100 ml of 10% solution), available from Sigma
11
,
or chemical equivalent. Makes a 0.1% solution. This should cost approximately
US$16 for 100 ml, or US$26 for 1 liter. Technical specification sheet and MSDS
are available at www.snowpure.com/knowledgebase.
Caustic Cleaning Solution Formula 1
(best, contains approved buffers and sequestering
agents)
:
9.8 liters deionized water
200 ml RoClean
tm
L211 solution (100 ml for mild fouling), available from Avista
Technologies. Technical specification sheet and MSDS are available at
www.snowpure.com/knowledgebase.
The pH will be 10.5-11.0.
The solution should be heated to 35-40°C for aggressive cleaning.
Caustic Cleaning Solution Formula 2:
9.8 liters deionized water
Sodium Hydroxide (pure NaOH) to make a 0.1-0.5% solution. Target pH 12-13.
The solution should be heated to 35-40°C for aggressive cleaning.
Procedure:
Do not “brine” the resins or the EDI product quality will never recover.
Plug the concentrate and electrode inlet and outlet ports. Isolate the
Feed
and
Product
ports from the RO and the product plumbing.
The cleaning of the resins is DOWNWARD FLOW.
Connect the chemical transfer pump to the
PRODUCT
port.
Connect the
FEED
port to a waste recovery tank, or for circulation connect it
to the chemical feed tank.
11
Triton X-100
tm
: CAS 9002-93-1, is also known as t-octylphenoxypolyethoxyethanol, or Octoxynol-9.