Lit. No. 76616, Rev. 02
25
November 15, 2018
FILL TANK SIZE:
Selecting the VOLUME position
on page 1 allows the
user to enter a FILL
TANK SIZE which
can be toggled to
OFF or 1 to 65535 using the
"+"
or
"
−
"
(Reset) keys. The FILL TANK
SIZE is in gallons in English or liters if
in Metric units. If a FILL TANK SIZE is
speci
fi
ed, then pressing
"+"
while set
to the
TANK
position (in Operation) will set the TANK
value to FILL TANK SIZE for quick reloading.
FLOW
CAL
VOLUME
(1) (2) (3)
TANK ALARM SET POINT:
Selecting the
VOLUME/MINUTE
position on page 1
allows a TANK
ALARM SET POINT to be entered.
When enabled, the Warning LED
fl
ashes, the alarm sounds, and the
display alternates between FILL and normal when
Tank Volume falls below the Set Point. When set to
"OFF," the alarm is disabled.
MIN
FLOW
VOLUME /
MINUTE
BLAST TARGET RATE:
Selecting the RATE position on page 1
allows the user to enter the desired
BLAST TARGET RATE. The units
and range are identical to the normal
TARGET RATE but
BLAST TARGET
RATE is only used
when the
BLAST
key is pressed.
BLAST DURATION:
Selecting the TANK LEVEL position
on page 1 allows the user to enter the
desired BLAST
DURATION. It
can be adjusted
from 0 (OFF) to
255 seconds.
TARGET
RATE
RATE
ADJUST
RATE
TANK LEVEL
"SPECIAL" CALIBRATION (CONT.)
Entering Calibration Values (cont.)
Special Calibration,
Page 1
UNITS (English or Metric):
Selecting the AREA
position on page 1
allows the user to
change the UNITS
and load defaults.
Pressing the
"+"
or
"
−
"
(Reset) keys
will toggle between English and Metric
UNITS and the display will show En9 or
nnEt.
NOTE: Trying to select Metric Units in Turf Mode will
cause the Error message to be displayed. Defaults will
not be loaded until the Rotary Selector is moved away
from AREA or the "Special" Cal page is changed or
"Special" Cal is exited normally.
WIDTH
CAL
AREA
(1) (2) (3)
VEHICLE NUMBER:
Selecting the
DISTANCE
position on page
one (1) allows the user to change the
VEHICLE NUMBER from 0 to 255 by
using the
"+"
or
"
−
"
(Reset) keys. The
VEHICLE NUMBER is downloaded with the Total and
Con
fi
guration data through the serial port.
SPEED
CAL
DISTANCE
(1) (2) (3)
VALVE POLARITY:
Selecting MODE position on page 1
allows the operator to install the
fl
ow
control valve in either a By-Pass or In-Line
con
fi
guration.
Pressing the
"+"
or
"
−
"
(Reset) key will
toggle the display between bYPAS and InLin.
MODE
(ALARM RESET)
MODE
SELECT
VALVE SPEED:
Selecting the SPEED position on
page 1 allows VALVE SPEED to be
changed, to adjust system response
speed (if required). Pressing the
"+"
or
"
−
"
(Reset) keys will adjust the VALVE SPEED
from –4 to 3. Normal setting is –1.
NOTE: Use caution
when adjusting the
Valve Speed setting.
Higher values used with a fast valve may cause
system instability (hunting, oscillations).
TEST
SPEED
SPEED
(1) (2) (3)
NOTE: Changing UNITS will load defaults, so they should
always be changed
fi
rst and then all other CALIBRATE
and "SPECIAL" CALIBRATE values may be set.
OPERATING AND CALIBRATION INSTRUCTIONS