Chapter 4: Alarms
4-4
Operation Manual SurgiVet
®
V3404 Plus
System Information Signal: Low Battery
A low battery condition will be detected when the battery has about 10 minutes remaining. As soon as the
condition is detected:
The message “
LOW BATTERY
” is displayed on the second message line. It alternates with other
messages at once per second.
This message remains displayed until the monitor is connected to power.
The low priority alarm LED is lit.
A unique tone, a burst of 5 beeps, sounds as soon as the low battery condition is detected, and every 30
seconds thereafter while the condition persists.
The volume of the low battery tone is equal to, or greater than, the alarm volume setting.
The low battery audible cannot be disabled by the
B
key.
Permanent Silence
When the monitor is turned on, simultaneously press the Wave/Trend
a
and Menu/Enter
b
keys to enter the
Service Mode. Use the ARROW (
on
) keys to toggle between
YES
and
NO
(enabled or disabled).
•
•
•
•
•
•