Advisor
®
Tech Vital Signs Monitor
7
Sp0
2
1.
Alarm Switch
: Adjustable (default = on)
2.
SpO
2
Alarm
: Select to change high and low values
3.
Sweep Speed
: Adjustable (default = 12.5 mm/s)
4.
Sweep Color
: Adjustable (default = blue)
ETCO
2
1.
Alarm Switch
: Adjustable (default = on)
2.
ETCO
2
Alarm
: Select to change high and low values
3.
AWRR Alarm
: Select to change high and low values
4.
Apnea Delay
: Setting range is 10 – 60 seconds (default = 10 seconds)
5.
ETCO
2
Unit
: mmHg, kPa or % (default = mmHg)
6.
Wave Scale
: 0 75 mmHg or 0 150 mmHg
7.
Fill In Waveform
: Adjustable
8.
Sweep Speed
: Adjustable (default = 12.5 mm/s)
9.
Sweep Color
: Adjustable (default = cyan)
10.
ETCO
2
Period
: 1 breath, 10 seconds or 20 seconds (default = 1 breath)
11.
Zero Setup
: Used to set the CO
2
module to zero
12.
Temperature Setup
can be found in the ETCO
2
menu where you are able to set
the following:
•
Alarm Switch
: Adjustable (default = on)
•
Temperature Unit
: Celsius or Fahrenheit (default = celsius)
•
Temperature Alarm
: Range for both T1 and T2 temperature measurement
channels is 25 – 50°C (upper limit default = 38°C, lower limit default = 36°C)
NIBP
1.
Alarm Switch
: Adjustable (default = on)
2.
Pressure Unit
: mmHg or kPa (default = mmHg)
3.
Patient Type
: Canine, feline, equine, bovine, ovine, caprine or swine (choose suitable
setting for your patient to avoid injury when the cuff inflates)
4.
Inflation Mode
: Manual, Auto and STAT (default = manual)
5.
Time Interval
: Used to support auto inflation mode, the range is 1 minute to 4 hours
What does the Apnea Delay
function do?
This setting is used to set the no
breaths detected time-out. This
time-out is the time period in seconds
following the last detected breath
at which the CO
2
module will signal
no breaths detected. The monitor
will alarm if the patient has stopped
breathing for longer than the pre-set
apnea time.
Helpful NIBP tip:
When in Auto Inflation Memory Mode,
ensure a new patient record has been
set up so that the inflation pressure
can reset itself and take a new unique
blood pressure measurement.