30
Annual Inspection and Testing
12. “Pump is starting…” should appear on the screen.
13. The main screen should appear with “Running” in the status bar, and the
green
LED indicator
light should blink every 3 seconds.
Note:
When the display is blank, the green, amber, or both LED indicators periodically
flash to indicate power and running status.
14. An alarm should sound, and “Reservoir volume low.” should appear on the display. Select
Acknowledge.
15. To stop the pump, press stop/start
9
. When the message “Stop pump?” appears, select
Yes.
16. The message “Pump is stopping…” appears, and the
amber
LED indicator light blinks.
“Stopped” should appear in the status bar.
☐
Activation Timing Check
17. Press stop/start
9
. The screen displays “Reset reservoir volume to 1 mL? Reservoir volume is
low.” Select
Yes.
18. The screen displays “Review pump settings.” Select
Review.
19. Confirm the settings match the parameters in the table in step 1.h. For each parameter select
Accept Value
and then select
Next.
20. The screen displays “Start pump?”. Select
Yes.
21. The display should show “Reservoir volume low.” Select
Acknowledge.
22. Start a timer at the first motor activation, and count the activations. One activation should
occur every 12 seconds. After approximately 1 minute, 50 seconds (1:50) and 10 activations, the
reservoir empty alarm should occur.
23. The display should show “Reservoir volume is zero. Pump stopped.” Select
Acknowledge.
☐
Remote Dose Cord Check
24. Attach the remote dose cord.
25. Program the pump with the following parameters:
Continuous Rate
PCA Dose
PCA Lockout
Reservoir
Volume
0.0 mL/hr
1.0 mL
1 min
10.0 mL
26. From the Tasks menu, select
View Reports.
27. Highlight and select the “Given and PCA Dose Counters” report.
28. Highlight “Total Given” and select
Clear Given.
29. Highlight “PCA Doses Given,” and select
Clear Doses.
30. Press stop/start
9
.
31. The screen displays “Review pump settings.” Select
Review.
32. Select
Accept Value
for all the parameters, then select
Next.
The screen displays “Start pump?”.
Select
Yes.
33. After “Running” appears on the status bar, press the remote dose cord button and note the
time. The pump should beep and begin to deliver.
34. Count the number of pump activations. The pump should make 10 double activations. After 10
double activations, the display should show a reservoir volume of 9.0 mL.
35. Press the remote dose cord button 2 more times within the next 1 minute. The pump should not
deliver, and the message “PCA dose not available. Currently locked out.” should display.