17
Pump Overview
Compatible Cassettes, Extension Sets, and Administration Sets
The cassette is the part of the CADD™ medication cassette reservoir or CADD
®
administration set that
attaches to the bottom of the pump. The following single-use products are compatible with the pump:
•
CADD™ medication cassette reservoir with a CADD
®
extension set
•
CADD
®
administration set
•
CADD
®
high volume administration set (with pump model 2110 only)
Notes:
•
A CADD
®
set with free-flow protection must be used in order to prevent free-flow.
•
The maximum delivery rate is 250 mL/hr on a model 2100 pump, and 500 mL/hr on a model
2110 pump. Programmed delivery rates exceeding 250 mL/hr require the use of a high volume
administration set and a model 2110 pump. Attempting to attach a high volume administration set
to a model 2100 pump will result in an alarm, and the pump will not run. Do not manually adjust
the delivery rate to override pump programming without written authorization from the prescribing
physician.
•
Smiths Medical recommends that the appropriate supplies needed to replace the cassettes are available
in case of a damaged cassette.
•
For detailed instructions and warnings pertaining to the CADD™ medication cassette reservoir or
CADD
®
administration set, refer to the instructions for use supplied with the product for preparing
the product for use.
Remote Dose Cord
Smiths Medical provides a remote dose cord for use by the patient. The push button is a
single pole double throw (SPDT) switch. When the remote dose cord is attached to the
pump, the patient may press the remote dose button to receive a PCA (PCEA) dose. For
easy access, the remote dose cord may be fastened to the patient’s clothing or bedsheet with
the attached clip.
Note:
To detach the remote dose cord from the pump, grasp the remote dose cord
connector and pull back using a straight, steady motion.
CAUTION:
Do not twist or turn the connector or use any instrument to remove it from
the pump.
For additional specifications, refer to the instructions for use provided with the product.
Pump Software Upgradeability using SureLink® Remote Support Software
The CADD®-Solis Ambulatory Infusion Pump software is upgradeable (when updates are available from
Smiths Medical) using SureLink® Remote Support Software – PC Direct Connect. This application,
available separately, may be used to receive available pump software updates from Smiths Medical
Customer Service via the Internet and install those updates onto eligible pumps via USB connection. For
more information about using SureLink® Remote Support Software, see the documentation provided with
the software.
If you have questions about pump software updates or SureLink® Remote Support Software, contact Smiths
Medical. When calling, specify the pump’s software version number. This information is located in the
pump’s Device Information Report (see the
Reports
section in the CADD
®
-Solis Operator’s Manual for