10011896-001
7
Installing New Batteries
If
LowBat
or
Battery Depleted
appears on the screen, you
should change the batteries.
Use two new, AA alkaline batteries such as the DURACELL® or
EVEREADY® ENERGIZER®.
Dispose of used batteries in an environmentally safe manner,
and according to any regulations which may apply.
WARNING:
•
Do not use rechargeable NiCd or nickel metal hydride
(NiMH) batteries. Do not use carbon zinc (“heavy
duty”) batteries. They do not provide sufficient power
for the pump to operate properly, which could result in
death or serious injury.
•
Always have new batteries available for replacement.
If power is lost, non‑delivery of medication will occur
and, depending on the type of medication being
administered, could result in death or serious injury.
•
If the pump is dropped or hit, the battery door or tabs
may break. Do not use the pump if the battery door
or tabs are damaged because the batteries will not be
properly secured; this may result in loss of power, non‑
delivery of medication, and, depending on the type
of medication being administered, death or serious
injury.